17
2021-01
9.
Technical data
Storage temperature:
-20 °C ... 70 °C
Operating temperature:
0 °C ... 40 °C
Dimensions:
200 x 100 x 50 mm
Weight:
450 g
Measuring principle:
Absorption measurement with single beam
photometer (Fig. 2), chopped operation
Projector:
LED
Spectral apparatus:
Interference filter
Measuring wavelengths:
365 nm and 520 nm
Spectral half-width value:
~ 5 nm
External light influence:
Negligible
Interface:
V24 (9600, 8, n, 2)
Power supply:
6 V ... 12 V DC
Current consumption:
max. 250 mA
Warm-up time:
0 min
Interference suppression:
According to DIN VDE 0871 and DIN VDE 0875
Inaccuracy:
< 0.5 % at A = 1.000
Relative photometric
short-time standard deviation:
< 0.1 %
Measuring ranges:
DP 310
DP 310 SI
CK in blood /
CK 321
0.0 - 2500 U/L
0.0 - 2500 U/L
CK in serum /
CK 121
0.0 - 2000 U/L
0.0 - 2000 U/L
GOT (ASAT)
10 - 500 U/L
10 - 500 U/L
GPT (ALAT)
10 - 500 U/L
10 - 500 U/L
Lactate
0.2 - 30 mmol/L
0.2 - 30 mmol/L
Lactate-rapid
0.2 - 20 mmol/L
0.2 - 20 mmol/L
Urea
5 - 200 mg/dL
0.8 - 35 mmol/L
Glucose
20 - 630 mg/dL
1.1 - 35 mmol/L
Triglycerides
20 - 2000 mg/dL
0.2 - 23 mmol/L
HDL-Cholesterol
10 - 200 mg/dL
0.2 - 5 mmol/L
Cholesterol
20 - 1300 mg/dL
0.5 - 35 mmol/L
Creatinine
0.0 - 5 mg/dL
0.0 - 440 µmol/L
Haemoglobin
(SLS-method)
0.0 - 50 g/dL
0.0 - 31 mmol/L
Erythrocytes
1.0 - 10 Mio/µL
1.0 - 10 Mio/µL
Haematocrit
5 - 90 %
5 - 90 %
ABS 365 nm
A = 2.500
A = 2.500
ABS 520 nm
A = 2.500
A = 2.500