Diaglobal DP 220 Скачать руководство пользователя страница 6

2021-01 

3.1

 

Power supply 

 

The  Gyn  Photometer  can  be  operated  as  desired  using  a  power  supply,  a  (9V 
block) battery or (model 6F22 or PP3) rechargeable battery. 

 
 
3.1.1

 

Mains power operation 

 

The  Photometer  is  supplied  with  a  power  supply  unit  for  operation  on  a  mains 
voltage  in  the  range  of  100  V  ...  240  V  AC.  The  mains  plug  is  marked  with  a 
Diaglobal logo (sticker). 
The  connector  plug  of  the  power  supply  unit  is  connected  to  the  power  supply 
socket on the back of the device. 

 
 
3.1.2

 

Network-independent operation 

 
To insert the rechargeable battery or the normal battery: 
Unscrew  the  knurled  screws  on  the  bottom  of  the  unit  and  remove  the  battery 
compartment  cover. Connect  the battery to the push-button contact and insert  it 
into the device. Replace the battery compartment cover and screw in the knurled 
screws. 
 
Please note: 
The  Gyn  Photometer  can  be  operated  using  a  power  supply  without  the  need  to 
remove the rechargeable battery or the normal battery. 
The  rechargeable  battery  cannot  be  charged  while  it  is  installed.  A  separate 
battery charger is required for this purpose. 

 
 
3.2

 

Measuring system 

 

The optical section is shown in Fig. 2. 
 

 

 

Fig. 2 
 
The  light  emitted  by  an  LED  is  first  selected  into  its  wavelength  ranges  (520  nm 
and  546  nm)  by  an  interference  filter  IF  (HBW  ~  5  nm)  and  then  bundled  and 
directed  onto  the  cuvette  in  the  shaft.  After  passing  through  the  cuvette,  a 
broadband  photosensor  converts  the  light  falling  on  its  sensor  surface  into  a 
current, proportional to the intensity. 

Содержание DP 220

Страница 1: ...1 2021 01 Gyn Photometer DP 220 Operating Manual Version 5 9 5 9 SI Edition 2021 01 ...

Страница 2: ...gher an automatic test of the device function has also been integrated Therefore the Gyn Photometer complies with the requirements of the guidelines of the German Medical Association With the Gyn Photometer 4 clinical chemical parameters can be determined The device can be supplied with SI units of measurement on request see chapter 9 Technical Data table Measuring Ranges The kits and accessories ...

Страница 3: ... Reagents parameter list 8 5 3 Control materials 8 5 4 Laboratory aids 8 6 Quality control according to the Guideline of the German Medical Association 9 7 Measuring procedure 9 7 1 End point measurement 9 7 2 Multipoint measurement taking the sample s blank value and calculation of the end point into account 10 8 Measurement 10 8 1 Switching the device on 10 8 2 Self test when switching on 10 8 3...

Страница 4: ...nd serum plasma The Gyn Photometer fulfils the basic requirements of Appendix I of Directive 98 79 EC regarding in vitro diagnostics The conformity of the device with Directive 98 79 EC is confirmed by the use of the CE marking Manufacturer Diaglobal GmbH Innovationspark Wuhlheide Köpenicker Str 325 Haus 41 12555 Berlin Tel 49 0 30 6576 2597 Fax 49 0 30 6576 2517 E Mail info diaglobal de http www ...

Страница 5: ...netic waves Operation on a horizontal surface Please observe the following instructions for use Insert a rechargeable battery or normal battery if the device is to be operated independently of a power supply or connect the photometer to a power supply unit Press the ON ENTER key Fig 1 to activate the internal device check which is automatically carried out by the device The device is then immediat...

Страница 6: ... compartment cover Connect the battery to the push button contact and insert it into the device Replace the battery compartment cover and screw in the knurled screws Please note The Gyn Photometer can be operated using a power supply without the need to remove the rechargeable battery or the normal battery The rechargeable battery cannot be charged while it is installed A separate battery charger ...

Страница 7: ...Mikrozid AF Liquid Bacillol plus 3 Kohrsolin or similar are recommended for cleaning the device and the surface Before cleaning the unit with a soft cloth and the decontaminating solution it must be switched off and the electrical power supply must be disconnected Make sure that no liquids get into the device There is no protection against penetrating liquids Code IP X0 The cuvette shaft must not ...

Страница 8: ...tocrit control Control blood for accuracy and precision control of haematocrit and erythrocytes determination in blood in the normal range 5 x 1 mL GLU QS Glucose control 100 mg dL 3 x 4 mL 5 4 Laboratory aids Art no Description Contents LH 001 Blood lancets 500 LH 004 Capillaries 10 µL with ringmark 250 LH 006 Cuvette rack 1 LH 007 Micropipettor pipetting aid 1 LH 009 Cellulose swabs 500 LH 010 C...

Страница 9: ...unts In agreement with the requirements of the German Medical Association a test of the device function see operating instructions chapter 8 5 is integrated in the Gyn Photometer therefor a daily test by means of a standard manual test German Medical Association part B chapter 2 1 5 paragraph 2 is not necessary The Gyn Photometer is suitable for the speedy detection of gestational diabetes and ful...

Страница 10: ...after it is switched on It takes approx 5 seconds after which the unit is ready for measuring Note If it becomes obvious during the test that one of the device functions does not correspond to the required settings SERVICE will appear in the display In this case switch the device off Please call Diaglobal GmbH service Tel 49 0 30 6576 2597 or contact your specialist retailer 8 3 Test selection Pre...

Страница 11: ... the individual measured values If specific requirements e g E2 E1 for ascending reactions are not met an error message is displayed 8 6 Notes on taking samples and carrying out measurements Errors in taking samples will always lead to incorrect measurement results This chapter addresses the most common errors that can occur during taking samples and measuring samples 1 Before measuring cuvettes s...

Страница 12: ... 1 mm from the ring mark is sufficient to obtain a completely incorrect measurement result If the sample is above the black ring mark this will lead to incorrect positive measurement results A cellulose swab can be used to carefully soak up too much blood If the sample is below the black ring mark this will lead to incorrect negative measurement results In this case correction is hardly possible d...

Страница 13: ... light influence Negligible Interface V24 9600 8 n 2 Power supply 6 V 12 V DC Current consumption max 250 mA Warm up time 0 min Interference suppression According to DIN VDE 0871 and DIN VDE 0875 Inaccuracy 0 5 at A 1 000 Relative photometric short time standard deviation 0 1 Measuring ranges DP 220 DP 220 SI Glucose 20 630 mg dL 1 1 35 mmol L Haemoglobin HiCN method 0 0 50 g dL 0 0 31 mmol L Haem...

Страница 14: ...irements Part 2 6 Particular requirements In vitro diagnostic IVD medical equipment 9 EN 592 Instructions for use for in vitro diagnostic instruments for professional use National directives and recommendations Germany 10 Guidelines for Quality Assurance of Laboratory Examinations of the German Medical Association of 23 12 2019 Note on electromagnetic compatibility a The photometer meets the requi...

Страница 15: ...ss ON ENTER key Wait for device check and confirm with ON ENTER 2 Select test Press arrow key until required test appears 3 Confirm required test Press ON ENTER 4 Switch off Press both arrow keys at the same time Note If SERVICE appears in the display after the device check the device has a defect In this case please contact our customer service at 49 0 30 6576 2597 ...

Страница 16: ...ro point adjustment Insert an unprocessed cuvette from package into photometer Zero point is stored in memory 4 Switch photometer on with ON ENTER key Wait for device check confirm with ON ENTER Select the required test confirm with ON ENTER In regard to series measurement After zero point setting any number of additional samples can be measured ON ENTER ON ENTER HB SLS Insert blank Zero HB SLS 0 ...

Страница 17: ...Zero point adjustment Insert cuvette with sample Fig 3 into photometer zero point is stored in memory Remove cuvette after signal tone 4 Switch photometer on with ON ENTER key Wait for device check confirm with ON ENTER Select the required test confirm with ON ENTER 6 Replace screw cap with starter cap ON ENTER ON ENTER M1 LAC Sample 01 insert 9 Time is displayed wait for measured value 7 Turn cuv...

Отзывы: