4.
Plug the control unit cable
into the sonication bath.
3.
Remove the rubber cap (upper left hole marked machine cable) from the back of the
soundproof box and feed the control unit cable through the hole. The fitting will secure
the cable to the sound-proof box. Turn the front section counterclockwise to seal.
Be sure there is enough slack in the cable.
6.
Place the sonication bath into
the soundproof box.
2.
Place sonication bath in
front of the soundproof
box.
5.
Place the motor lid on the top of the sonication bath and
connect it. Make sure that the cable of the motor lid is
connected as shown in the image (triangle have to match).
8.
Connect the power supply adapter to the control unit
with the power cable. (Power supply adapter is only
needed outside the U.S.)
9.
Place the control unit on
the top of the soundproof
box.
Plug the power cable into
the outlet and switch on the
power switch on the back
of the control unit.
Now you are ready to start !
7.
Fill the sonication bath up to the red fill line with
distilled water only. Do not use deionized water!
1.
Open the boxes and unpack all components.
Installing the Bioruptor
®