Input Limits
Never apply voltage or current to the meter that exceeds
the specified maximum:
Function
Maximum
V DC or V AC
1000VDC/AC rms
mA DC/AC
500mA 1000V fast acting fuse
A DC/AC
10A 1000V fast acting fuse (20A for
30 seconds max every 15 minutes)
Frequency, Resistance,
1000VDC/AC rms
Capacitance,
Diode test, Continuity
Temperature
1000VDC/AC rms
Surge Protection: 8kV peak per IEC 61010
1.
Use extreme caution
when working with high
voltages.
2.
Do not
measure voltage if the voltage on the
“COM” input jack exceeds 600V above earth
ground.
3.
Never
connect the meter leads across a voltage
source while the function switch is in the current,
resistance, or diode mode. Doing so can damage
the meter.
4.
Always
discharge filter capacitors in power sup-
plies and disconnect the power when making
resistance or diode tests.