FPG8601™/VLPC™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 128
8.5
MOUNTING POST PRT CALIBRATION
The temperature as read by the mounting post PRTs is used to determine the thermal expansion of the
piston-cylinder effective area. The temperature is also used to monitor temperature rate of change and
temperature change from the time of zero. Accuracy of these PRTs is critical to proper FPG operation.
This section describes how to remove and calibrate these sensors.
The temperature sensor inside the lubricating volume of the FPG is used only for an indication of
temperature change. The value is not used for FPG pressure determination. Therefore, this sensor is
not directly calibrated.
8.5.1
REMOVE THE PRT
S
Make sure that the FPG is vented in gauge mode. Then turn the power off.
Detach the PRT cable from the FPG platform by unscrewing the connection in a counter
clockwise direction.
Remove each PRT from the mounting post by unscrewing the knurled nut from each
mounting post PRT. Note that there are two o-rings that hold each PRT into the
mounting post. Remove and save these o-rings.
Remove the knurled nut from each PRT.
8.5.2
CREATE THE PRT CABLE ADAPTER
Refer to Figure 66 to create an adaptor cable to calibrate the PRT.
Figure 66.
PRT Calibration Cable Adaptor
Table 47.
PRT Calibration Cable Connections
FROM
TO
SIGNAL
P1-1 E1
I
Source
P1-2 E3
Upper
(,1)
+
P1-3
E4
Upper (,1) -
P1-4 E2 I
Ret
P1-5 E5
Lower
(,2)
+
P1-6
E6
Lower (,2) -
Содержание FPG8601
Страница 1: ...2007 DH Instruments a Fluke Company FPG8601 VLPC Operation and Maintenance Manual...
Страница 10: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page X N NO OT TE ES S...
Страница 12: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page XII N NO OT TE ES S...
Страница 22: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 10 N NO OT TE ES S...
Страница 48: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 36 N NO OT TE ES S...
Страница 118: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 106 N NO OT TE ES S...
Страница 144: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...
Страница 157: ...11 APPENDIX Page 145 2007 DH Instruments a Fluke Company 11 3 5 2 ABSOLUTE PRESSURE...
Страница 160: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 148 11 5 1 SYSTEM LEVEL GROUNDING...
Страница 161: ...11 APPENDIX Page 149 2007 DH Instruments a Fluke Company 11 5 2 FPG GROUND DIAGRAM...
Страница 162: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 150 11 5 3 VLPC POWER DIAGRAM...
Страница 164: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 152 N NO OT TE ES S...
Страница 166: ...FPG8601 VLPC OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S...