3. OPERATION
Table 4.
AMH Errors (continued)
AMH
ERROR
NUMBER
PG TERMINAL MESSAGE
COMMENT
120
<AMH unable to lift mass load>
Mass lift valve not sensed.
121
<AMH unable to lower mass load>
Mass is dropping unexpectedly.
122
<AMH unable to lift mass load>
Mass never lifted.
123
<AMH unable to lower mass load>
Mass is not dropping when it should.
124
<AMH unable to lift mass load>
IR beam malfunction. AMH will not attempt lift
125
<AMH unable to rotate column
motor>
Column motor switch on column 1 is misaligned.
126
<AMH unable to rotate column
motor>
Column motor switch on column 2 is misaligned.
130
<AMH valve failure>
6.2 kg binary mass pin engage valve (VLV#14) not sensed.
131
<AMH valve failure>
3.2 kg binary mass pin engage valve (VLV#6) not sensed.
132
<AMH valve failure>
1.6 kg binary mass pin engage valve (VLV#12) not sensed.
133
<AMH valve failure>
0.8 kg binary mass pin engage valve (VLV#4) not sensed.
134
<AMH valve failure>
0.4 kg binary mass pin engage valve (VLV#10) not sensed.
135
<AMH valve failure>
0.2 kg binary mass pin engage valve (VLV#2) not sensed.
136
<AMH valve failure>
0.1 kg binary mass pin engage valve (VLV#8) not sensed.
137
<AMH valve failure>
6.4 kg binary mass pin disengage valve (VLV#15) not
sensed.
138
<AMH valve failure>
3.2 kg binary mass pin disengage valve (VLV#7) not sensed.
139
<AMH valve failure>
1.6 kg binary mass pin disengage valve (VLV#13) not
sensed.
140
<AMH valve failure>
0.8 kg binary mass pin disengage valve (VLV#5) not sensed.
141
<AMH valve failure>
0.4 kg binary mass pin disengage valve (VLV#11) not
sensed.
142
<AMH valve failure>
0.2 kg binary mass pin disengage valve (VLV#3) not sensed.
143
<AMH valve failure>
0.1 kg binary mass pin disengage valve (VLV#9) not sensed.
Page 29