Hardness Tester DHT-100
PLUS
18
2. Click this new disk, you will find all stored data is listed as F001, F002 in
the file “MD_ARY”. Users could open this file and check the data, or even copy
them to any software for further analysis.
6.7 Deleting of wrong measuring data
During measurements, some measuring values are obviously error data;
press the key
to delete these values conveniently.
7. Repairing and Maintenance
In order to keep the accuracy and reliability of the instrument, it is
necessary of timely evaluation and maintenance.
7.1 Maintenance of battery
DHT-100Plus works with 2PCS of AA alkaline battery. It is easy for users to
replace if it is in low battery condition.
Notice: If this main unit is not used for so long, please take out the
batteries for free of corrosion to the electronic board.
7.2 Maintenance of impact body tip
Owing to the abrasion of impact body tip,
if it shows within the range of
tolerance (±6HLD) on testing the block comes with the delivery, but out of
range of repeatability (the maximum value minus the minimum value less
than 12HLD), please replace the impact body and do calibration again. If it
does not show within the range of tolerance, but repeatability is good, please
make change through the Menu CALIBRATION. For details please refer to
“6.5.3 calibration”.
7.3 Maintenance of impact device cable
During measurements, the impact cable may be damaged, just replace it
with a new one.
7.4 Maintenance of printer
If print data is not clear, there must be dust inside the printer. Please open
the shell of printer, take out the paper, and then clear inside with close or
paper. After that, print data will be clear again.