SI units, as selected in the <Unit> area (see above).
Mark Positive:
Allows you to choose whether the positive results are
highlighted in red color. (
Yes/No
)
Select <Yes> to make a distinction between positive(red) and normal(blue)
results.
Select <No> if you wish to express all results in the same color.
4.6.4.2
System Setting
This allows selecting the measurement Speed, Sound
and Baud rate.
[Fig. 1.3.9] System Setting
Speed:
To select the measurement speed. Default speed is
<General>.
The <
General
> mode includes an incubation time inside the instrument. So
after handling, you can place the strip on the loader and the strip will be
read at the correct time.
<
General
>mode = 60seconds of operation by the instrument
By using the <
Quick
> mode, the analyzer immediately reads the strip
without the incubation time. This implies that the soaked strip should be
incubated at room temperature for 30 seconds prior to launching
analysis(=outside of the reader).