4
and the resulting corrected chloride value is multiplied by the size and extraction multipliers and
divided by the chloride fraction (percent chlorine divided by 100):
[Analyte] = ([Cl
-
]
raw
- [Cl
-
]
blank
)(size multiplier)(extraction multiplier)/(chlorine fraction)
The
L2000
DXT
Analyzer comes preprogrammed with 33 methods providing for analysis of a
large number of organo-chlorine compounds in most matrices. Correction factors are
preprogrammed to account for all of the steps necessary to run samples using the standard
prepackaged reagents. If the procedures are modified or other analytes are expected, up to 20
custom methods can easily be built.
The operational program of the
L2000
DXT
has been designed to provide the most versatile data
reduction platform possible for the analysis of chlorinated organic compounds in a variety of
matrices. At the same time, the program is easy to use for routine analysis and does not require
custom method development.
Unpacking
The
L2000
DXT
is shipped in its own carrying case, complete with all the hardware necessary for
operation. Upon receiving the
L2000
DXT
Analyzer, please verify that all of the items listed
below are present and in good working order.
The carrying case for the
L2000
DXT
Analyzer should contain the following items:
L2000
DXT
Electronic PCB/Chloride Analyzer
Power cube AC-DC transformer
Portable electronic balance & 100 g Calibration weight
5 mL pipettor
Vial rack
Timer
Marking pen
USB thumb drive
Two empty 20 mL glass vials for RINSE and CALIBRATION solutions
User’s Manual, Certificate of Calibration and MSDS
NOTE: DO NOT return meter to case with USB Drive installed.
In addition to the aforementioned items in the carrying case, the following should be included in
the same outer shipping box:
Chloride-ion specific electrode
1
Packet of polishing strips
1
Test tube rack
To prevent damage to the instrument in the case of a leak, the sample preparation reagents are
shipped separately. The exact makeup of the reagent pack will depend on the specific reagent
1
These items are shipped in their original packaging from Orion. Once they have been unpacked, they should be carried in the right-hand slot in