TRC:
After selecting the TRC parameter, press
or
to select the sampling mode (AVG or PEAK),
AVG: sample average mode, the display result of each
point will be an average of several sample value. It will
be faster than the PEAK mode, and also if you want to
measure the noise of the system, you need to select
this mode.
PEAK: positive-peak mode, the display result of each
point will be peak value of several sample value. It will
be necessary to select this mode to measure the video
or audio level of analogue channels. Show as figure
3-6-4.
Figure 3-6-4