11. Remove the grommet that seals the blower’s wire harness.
(Fig. 7)
12. Remove the heater wire harness connector from the bot-
tom of the chassis by sliding the connector outward while
pulling the top retaining tab upward slightly. (Fig. 8)
13. Remove the 1/4” blower wrap foam, the blower, and the
silicone isolator from the underside of the chassis.
14. Carefully spray, wipe or soak the air intake, fl ow path, and
blower chassis inside and outside using a 2% glutaralde-
hyde solution.
15. Wipe the power cord and the outside of both halves of the
PAP case with the same solution.
16. Let all components fully dry
17. Reassemble the unit using parts from the Disinfection
Kit—DV54/DV53D-682.
NOTES:
• Route the silicone tubing correctly, as shown, and ensure
that there are no kinks. (Fig. 9)
• Ensure that the silicone isolator is fully seated in the chas-
sis and the blower.
• Ensure that the grommet is fully seated around the blower
wire harness.
• Route the wire harnesses properly, using the provided
strain reliefs on the chassis. (Fig. 10 & 11)
• Secure the power supply board within the mount clips on
the chassis before replacing the unit covers. (Fig. 10)