Sentinel CX
Sentinel CX Sensor Station IM
Rev. 1.5
Page 18 of 26
4. Recharging the Internal Battery Pack
4.1 Non-Hazardous and Indoor Location
At a safe time interval prior to full discharge of the internal battery pack, relocate the CX Sensor Station to
a non-hazardous indoor location area that has suitable 110 VAC or 220 VAC AC power receptacles
available.
4.2 Connecting to Mains Supply
Using the appropriate Battery Charging Accessory product (PN
976-0003BC-00T
for 110 VAC or PN
976-
0003BC-220
for 220 VAC), find the corresponding Mains Supply receptacle in an indoor non-hazardous
area and plug-in the pronged end of the cable. Connect the Turck connector to the external port labeled
external power. Turn on the power switch and begin charging. Operate the charger for the necessary time
it takes to accomplish a full charge.
CAUTION
Only the Mains supply cord that is supplied with the Detcon-approved Battery
Charger Accessory product shall be used. Use of any other Mains supply cords will
impair the safety of this device.
Warning
Charging of the internal Battery Pack shall never be done while equipment is
located in the Division 2 hazardous area. The equipment must be moved to a non-
hazardous indoor location area for the battery charging to take place.
The
temperature during the indoor battery charging shall be between 0-40C.
Warning
The charging power supply is only safely rated for indoor use, so use of the Detcon
charging accessory is limited to indoor non-hazardous areas.