COM Mode

Operation

Using the Meter

1. Press the EXIT button to power the meter.
2. Hold the test probe and momentarily touch the

ground touch plate located on the back of the unit.

Figure 4. Momentarily touching the ground touch plate to 

ground the operator 

Exports data to PC using supplied 8-Pin DIN to USB 

cable.

Installation

1. Remove the meter from the carton.
2. Connect the included ground cord to one of the

two ground jacks located at the top of the meter.

Connect the opposite end of the ground cord to

equipment ground.

3. Connect the test probe to the probe connector

located at the top of the meter.

4. Press the EXIT button to power the meter.
5. NOTE: The electronic components of the

electrostatic sensor require a warm-up period

of 5 minutes to obtain accurate and stable

measurements. The sensor uses a vibrating

element to perform measurements, and the

warm-up period allows its mechanical assembly

to stabilize. An exclamation mark symbol on the

display and an audible sound will alert the operator

once warm-up has been completed.

Figure 3. Setup of the Body Voltage Meter

3. Verify the Body Voltage Meter is zeroed by toggling

to the MONITOR menu and reading the measured

value. It should be 0 V. Zero the Meter if required by

toggling to the ZERO menu.

Monitor Mode

1. Toggle to MONITOR mode on the meter, and press

the Start / Stop button.

2. The display will show the electrical potential present

on the probe.

3. A beeping sound (every second) indicates that the

meter is measuring. This is to ensure the unit is not

left in measuring mode unintentionally and cause

the battery to drain.

4. To stop the measuring process, press the Start /

Stop or Exit buttons.

6. The meter is now ready for use.

Test Equipment Depot - 800.517.8431 - 

5 Commonwealth Ave,

 Woburn, 

MA 01801 

 - TestEquipmentDepot.com 

Содержание TB-3081

Страница 1: ...verifications shall occur The compliance verification plan shall document the test methods used for process monitoring and measurements If the organization uses different test methods to replace those of this standard the organization shall be able to show that the results achieved correlate with the referenced standards Where test methods are devised for testing items not covered in this standard...

Страница 2: ...nto a PC H Ground Jack Common point ground for the meter Connect the included black ground cord here I Ground Touch Plate Touch to ground the operator prior to taking body voltage measurements Settings Mode Press and hold the Settings button for 1 second to enter the SETTINGS mode There are 4 menus available 1 V RANGE 2 CLOCK 3 DISPLAY 4 C F Press the Left button to move to the previous menu Press...

Страница 3: ...top button to toggle the setting Operation Modes There are 6 operation modes available 1 MONITOR 2 ZERO 3 NEW 4 VIEW 5 CLEAR 6 COM Press the Left button to move to the previous mode Press the Right button to move to the next mode Press the Start Stop button to enter the selected mode Press the Exit button to exit the selected mode Allows on the go measurements without saving any data Zero Mode Zer...

Страница 4: ...ccurate and stable measurements The sensor uses a vibrating element to perform measurements and the warm up period allows its mechanical assembly to stabilize An exclamation mark symbol on the display and an audible sound will alert the operator once warm up has been completed Figure 3 Setup of the Body Voltage Meter 3 Verify the Body Voltage Meter is zeroed by toggling to the MONITOR menu and rea...

Страница 5: ...Voltage Meter to a PC using the included Black 8 Pin DIN to USB Cable 7 Toggle to COM mode on the Body Voltage Meter 8 Run WTManager1 3 3 9 Click on Datalogger 10 Click on Com Port select 11 Click on Port The COM selection for the Body Voltage Monitor will become visible 12 When selected the unit will be highlighted 13 Click OK Figure 5 Using the Body Voltage Monitor 4 To stop the test press the S...

Страница 6: ... 1 4 x 5 1 Ø35 mm x 130 mm Weight meter with batteries 1 4 lbs 650 g Weight probe 0 6 lbs 0 27 kg Country of Origin Italy Importing Data into a PC 1 Click on the Datalogger option 2 Select Import Graph A progess bar will appear at the bottom of the screen 3 Data can either be saved or exported in Excel format Figure 9 Replacing the Batteries on the Body Voltage Meter Figure 8 Viewing the body volt...

Отзывы: