G. Battery Compartment:
Open this compartment
to install the four AA alkaline batteries needed to
power the meter. Replace the batteries once the
battery icon on the display is empty.
Home / Test Results Screen
Figure 5. Home/Test Results screen
H. Battery Life Indicator:
Displays the approximate
life of the meter's 4 AA alkaline batteries.
I. Relative Humidity:
Displays the relative humidity.
J. Test Voltage:
Displays the test voltage used to
complete the measurement.
K. Temperature:
Displays the ambient temperature.
L. Resistance Measurement:
Displays the
resistance measurement in ohms (
Ω
).
Settings Manu
Figure 6. Settings menu
M. Firmware Revision:
Displays the meter's
firmware revision.
N. Temperature:
Sets the unit of measurement for
temperature to either Fahrenheit (
º
F) or Celsius (
º
C).
O. Stabilization Mode:
Sets the meter's
Electrification period setting to either Auto and Fixed
Stabilization.
SCS
- 926 JR Industrial Drive, Sanford, NC 27332
East: (919) 718-0000 | West: (909) 627-9634
•
Website:
StaticControl.com
TB-9092
Page 4 of 8 December 2019
©
2019 DESCO INDUSTRIES INC
Employee Owned
Auto - Enables a 15-second electrification period
when the measured resistance is 1 x 10
10
ohms or
greater to maintain test accuracy.
Fixed - Complies with ANSI/ESD S4.1 and enables
A 15-second electrification period when the
measured resistance is 1 x 10
6
ohms or greater.
P. Beep:
Enables and disables the audible beep
when the meter's pushbuttons are pressed.
Q. Erase all memory:
Erases all stored
measurement transactions saved in the meter's
memory.
Memory Recall Menu
Figure 7. Settings menu
R. Memory Slot Number:
Indicates the memory
slot number.
S. Resistance Measurement:
Indicates the
resistance measurement value for the respective
memory slot.
Operation
General Guidelines
Use both of Single and Dual Surface
Electrodes for Resistance Point-to-Point (Rtt)
measurements.
Use the Single-Surface-Electrode, and connect the
black test lead to ground for Resistance-to-Ground
(Rtg) measurements.
Ensure that the item being measured is electrically
isolated (placed on an insulative surface). The
meter will measure the lowest resistance path.
Minimize crossing the test leads when possible.