3651 WALNUT AVENUE, CHINO CA 91710 • PHONE (909) 627-8178 •
Page
7
of
10
TB-3089 April 2021
© 2021 DESCO INDUSTRIES INC
Employee Owned
4. Point the plate in the airflow of
the ionizer to be measured and
press the START button.
Voltage will apply onto the plate, and the
meter will measure the amount of time
needed to neutralize it.
NOTE
•
If the display panel blinks "+1" or "-1",
the charged voltage has exceeded the
measurable range (over-range). If this
occurs, stop measuring immediately,
as this may cause damage to the
meter.
5. The meter will chirp and display
the neutralization time when the
measurement is completed.
The neutralization time in seconds is
displayed at the bottom of the display.
6. To turn the meter off, slide the power switch to the OFF
position.