3651 WALNUT AVENUE, CHINO CA 91710 • PHONE (909) 627-8178 •
Page
5
of
10
TB-3089 April 2021
© 2021 DESCO INDUSTRIES INC
Employee Owned
Operation
Measuring Ion Balance
1. Turn the meter on.
Slide the power switch to the ON position.
The meter will chirp then activate.
2. Zero point adjust the meter.
Bring the plate into contact with a
grounded body, and press the 0ADJ
button.
NOTE
•
Zero point adjustment must be reset
every time the meter is powered.
3. Set the desired mode.
Press the SELECT button, and the meter
will scroll through the modes shown in the
table on the right.
To measure balance (offset voltage), use
I.B mode.
4. Point the plate at the
measurement location.
The meter will measure the balance.
NOTE
•
If the display panel blinks "+1" or "-
1", the charged voltage has
exceeded the measurable range
(over-range). If this occurs, stop
measuring immediately, as this may
cause damage to the meter.
5. To turn the meter off, slide the power switch to the OFF
position.
Mode
LED display
I.B Mode
I.B
+DECAY Mode
+DECAY
-DECAY Mode
-DECAY