background image

TB-3048

 March 2011 Page 1 of 5

Made in America

© 2010 DESCO INDUSTRIES INC.

Employee Owned

Digital Static Locator  
Installation, Operation, and Maintenance

Description

Desco’s Digital Static Locator is a compact 

electrostatic fieldmeter used for locating and measuring 

static charges. Its pocket size makes it handy to use 

and four button operation makes it easy to operate.

The Digital Static Locator measures static voltages 

within ±20kV (20,000V) at a distance of 1". Results 

are simultaneously displayed numerically and in bar 

graph format. POWER on/off, ZERO adjustment, Ion 

Balance (IB) and HOLD are all push button operation. 

The HOLD button allows the display to retain the static 

charge reading.  This is especially useful where display 

is difficult to see during measurement.

Two LED guide ring lights help position the fieldmeter at 

the right distance from a charged test object. The 

conductive case and ground snap facilitate grounding

for accurate measurement. The circuitry of the 

19430

 

has been designed to make measurements in areas 

using air ionization.

Features

• Easy to read, easy to use operations

• Lightweight, compact design

• Range light assures accurate and repeatable 

  measurements

• Automatic range switching up to ±20kV (20,000V) at 1" 

  (25mm)

• Digital and bar graph display

• Automatic power off after five minutes (may be 

  disabled if preferred)

Figure 1. Digital Static Locator Item #

19430 

Specifications

Measuring range:

      0 to ±1.49 kV (Lo range)

 

                       ± 1.0 kV to ± 20.0 kV (Hi range)

 

                       (Range hysteresis: ± 1.0 kV to 

 

 

          ± 1.5 kV)

 

 

          0 to ± 200 V (Ion Balance 

 

 

          measuring range)

Measuring distance:

 1"±1/64" (25mm ± 0.5mm); LEDs  

 

 

          guide for correct distance 

 

 

          (between charged object and  

 

 

          fieldmeter)

Response time:

         <1 s

LCD display

 

renewal rate:

            5 times/s

Accuracy:

  

          20KV ± 10 %

Ambient conditions:

 50º-100ºF, (10º - 40ºC); 0 - 60  

 

 

          %RH (non-condensing)

Display features:      

 Bar graph; red LCD positive 

 

 

          polarity voltage, blue LCD

 

   

          negative polarity voltage

Bar graph 
precision:

  

          ±0.1 kV for low range ±1.5 kV for  

 

 

          high (Hi) range; ±15 V for ion 

 

 

          balance range

Digital reading:

          Auto-ranging, three digits . = 0 to  

 

 

          ±1.49 kV . = ±1.0 kV to 

 

 

          ±20.0kV (Hi) = 0 to 200 (Ion 

 

 

          Balance)

Ion Balance mode display: [IB] is displayed

Battery capacity display indicator

HOLD switch to retain display after measurement

Error sign [Error] appears if the sensor is damaged

Alarm feature:

             Beep sound will be heard during  

 

 

           the following actions:

Power on: one beep
Power on with Auto Off feature disabled: [Depress

power button for more than 3 seconds] three beeps
Auto power-off: short beeps at 1 second interval for 5

seconds before power off
Over range: Continuous sound

TECHNICAL BULLETIN TB-3048

DESCO WEST

 - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449 

DESCO EAST

 - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: 

Desco.com

Содержание 19430

Страница 1: ...splay Automatic power off after five minutes may be disabled if preferred Figure 1 Digital Static Locator Item 19430 Specifications Measuring range 0 to 1 49 kV Lo range 1 0 kV to 20 0 kV Hi range Range hysteresis 1 0 kV to 1 5 kV 0 to 200 V Ion Balance measuring range Measuring distance 1 1 64 25mm 0 5mm LEDs guide for correct distance between charged object and fieldmeter Response time 1 s LCD d...

Страница 2: ... TB 3048 Page 2 of 5 DESCO WEST 3651 Walnut Avenue Chino CA 91710 909 627 8178 Fax 909 627 7449 DESCO EAST One Colgate Way Canton MA 02021 1407 781 821 8370 Fax 781 575 0172 Web Site Desco com circuit This terminal must be grounded properly for accurate measurement 1 Turn the focusing LED and sensor toward an un charged area 2 Press the POWER button A single beep for one second indicates proper op...

Страница 3: ...nd to check for residual charges on surfaces The Digital Static Locator can also be used to measure ion balance voltage residual voltage 1 Depress the charge plate release button on either side of the fieldmeter to remove the charge plate assembly 2 Insert the plate assembly into the other end of the fieldmeter the end with the Desco logo until it clicks into place NOTE Ion Balance plate must not ...

Страница 4: ...he sensitivity of measurement of the instrument 3 Do not attempt to take apart the fieldmeter 4 The measuring distance between the fieldmeter and a test object should be 1 1 64 25mm 1mm The calibration was carried out at this distance The specified accuracy is valid for this distance Troubleshooting Guide Fault Possible Causes Solution Fieldmeter is on but when turned toward a charges object the n...

Страница 5: ...ce or other proof of date of purchase Any unit under warranty should be shipped prepaid to the Desco factory Warranty replacements will take approximately two weeks If your unit is out of warranty call our Customer Service Department at 909 664 9980 Chino CA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your ...

Отзывы: