TB-3060

 Page 5 of 5

© 2018 DESCO INDUSTRIES INC

Employee Owned

DESCO WEST 

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178

DESCO EAST 

- One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Website: 

Desco.com

Limited Warranty, Warranty Exclusions, Limit 

of Liability and RMA Request Instructions

See the Desco Warranty - 

Desco.com/Limited-Warranty.aspx

Calibration

Frequency of recalibration should be based on the critical 

nature of those ESD sensitive items handled and the risk of 

failure for the ESD protective equipment and materials. In 

general, Desco recommends that calibration be performed 

annually.
Use the Desco 

98221

 Wave Distortion Monitor Verification 

Tester to perform periodic verification (once every 6-12 

months) of the Full-Time Continuous Monitor. The verification 

tester can be used to check the test limits of the Full-Time 

Continuous Monitor without removing it from the workstation.
See 

TB-3074

 for more information.

Figure 7. Desco 

98221

 Wave Distortion Monitor 

Verification Tester

Specifications

Input Voltage and 

Frequency

100-120VAC, 50/60 Hz

Operating Temperature

50 to 95°F (10 to 35°C)

Environmental 

Requirements

Indoor use only at altitudes less 

than 6500 ft. (2 km)
Maximum relative humidity 

of 80% up to 85°F (30°C) 

decreasing linearly to 50% @ 

85°F (30°C)

Dimensions

4.38" L x 3.25" W x 0.89" H

(111 mm x 83 mm x 23 mm)

Weight

0.26 lbs. (0.12 kg)

Operator Test Range*

500 kilohms to 10 megohms 

impedance

Worksurface Test Limit** 10 megohms (±10%)
Operator Test Voltage

1.2V peak-to-peak @ 1.2µA, 

open circuit

Worksurface Test 

Voltage

5 to 7.5V, open circuit

Alarm Response Time

<50 ms

Country of Origin

United States of America

*This cannot be verified with standard DC test equipment. 

The continuous monitor is an impedance sensing device, and 

the limits are determined by the magnitude and angle of the 

impedance.
**The worksurface mat must have a conductive layer such 

as Dual Layer Rubber or Dissipative 3-Layer Vinyl with 

conductive buried layers. Desco workstation monitors are not 

recommended for use with homogeneous matting.

Содержание 19325

Страница 1: ...can save a significant amount of money in testing costs and rejected products Periodic testing detects failures after ESD susceptible products have been manufactured The costs of dealing with the resulting catastrophic or latent defects can be considerable Full Time Continuous Monitors eliminate the need for users to test wrist straps and log the results by their function these monitors satisfy th...

Страница 2: ...nitoring For example during each 2 2 second pulse cycle of a leading constant resistive monitor electrical current is pulsed for only 0 2 seconds followed by an unmonitored interval of 2 seconds This leaves the user wrist strap unmonitored for over 90 of each cycle Damaging static charges can easily occur in the portion of the time in between the pulses The off period of 2 seconds equals 2 billion...

Страница 3: ... terminal block on the monitor 3 Connect the tinned wire termination of the Mat Monitor Cord to the MAT terminal block on the monitor 4 Determine the mounting location of the Full Time Continuous Monitor It is normally installed toward the front edge of a workstation where its LEDs can be visible to the operator Use the included hook and loop strip if desired 5 Use a blade to pierce two holes 12 t...

Страница 4: ...itor s operator jack will disable the audible alarm PARKING SNAP All Desco single wire monitors with wave distortion technology include the Park Snap feature Use the Park Snap to store the operator s wrist cord and enable standby mode while the operator is away from the workstation Before disconnecting the coil cord from the wristband the operator touches the Park Snap to disable the audio alarm f...

Страница 5: ... more information Figure 7 Desco 98221 Wave Distortion Monitor Verification Tester Specifications Input Voltage and Frequency 100 120VAC 50 60 Hz Operating Temperature 50 to 95 F 10 to 35 C Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C decreasing linearly to 50 85 F 30 C Dimensions 4 38 L x 3 25 W x 0 89 H 111 mm x 83...

Страница 6: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Desco 19325 ...

Отзывы: