Maintenance

The Digital Surface Resistance Meter requires little 

maintenance. There are no user serviceable parts. If the 

meter requires service beyond cleaning the electrodes 

or replacing the batteries, please contact 

Desco Europe 

Customer Service

.

Battery Replacement

Replace the batteries once the battery life indicator 

is empty. Open the compartment located on the back 

of the meter to replace the batteries. The meter uses 

four AA alkaline batteries. Ensure that the batteries’ 

polarities are oriented in the correct fashion to avoid any 

possible circuit damage.

Cleaning the Digital Surface Resistance Meter

The area surrounding the test jacks at the top end of the 

meter should be wiped with a clean, isopropanol-alcohol 

moistened cloth to remove skin oils that will accumulate 

and affect the meter’s accuracy at high resistances. 

The frequency of cleaning will depend on usage. Desco 

Europe recommends cleaning this area once a month. 

Cable jackets should also be cleaned in this fashion.

Cleaning the 2.27 kg Electrodes

Clean the electrodes with a minimum 70 % isopropanol-water 

solution. Make sure the 2.27 kg electrodes’ conductive pads 

are dry prior to use.

Figure 13. Measuring the Resistance Point-to-Point (Rtt) 

of floor materials

Measure Resistance Point-to-Point (Rtt) on the 

Surface

1.  Do not clean the surface. 
2.  Remove from the surface only those items that 

might interfere with the test. ESD sensitive devices 

shall also be removed. 

3.  Connect both test leads to both 2.27 kg electrodes, 

and place the electrodes on the most commonly 

used area of the surface. Use the 25 cm electrode 

spacer to space them 25 cm apart from each other. 

The electrodes should also be 5 cm away from any 

edge and 8 cm away from any grounded point. If the 

most used area is not obvious, use two points near 

the center of the surface. 

4.  Push the red pushbutton to perform a measurement. 

Should the 15-second electrification period appear, 

it may be bypassed by pushing the red pushbutton 

a second time. 

Figure 12. Electrification period timer

5.  Push the right black pushbutton to save the 

measurement if desired. 

6.  Perform additional measurements by placing the 

electrode on the most commonly used or worn area.

Figure 12. Measuring the Resistance Point-to-Point (Rtt) 

of a floor tile

Содержание 19290

Страница 1: ...ooring system IEC 61340 4 5 A compliance verification plan shall be established to ensure the organization s fulfilment of the requirements of the plan Process monitoring measurements shall be conducted in accordance with a compliance verification plan that identifies the technical requirements to be verified the measurement limits and the frequency at which those verifications shall occur The com...

Страница 2: ...asurement Location Labels 25 Pack 222634 Replacement 2 27 kg Electrodes 50004 Replacement T Handles for 2 27 kg Electrodes 222002 Concentric Ring Probe Packaging 19290 Digital Surface Resistance Meter Kit 1 Digital Surface Resistance Meter 2 Test Leads 1 5 m Length 2 2 27 kg Electrodes 1 Electrode Spacer 25 cm 1 Electrode Spacer 91 cm 4 AA Alkaline Batteries 1 Ground Plug Adapter 1 Gator Clip 25 M...

Страница 3: ...ty battery life test voltage and resistance measurement D Power Switch Slide the switch to the left to power the meter OFF Slide the switch to the right to power the meter ON E Black Pushbuttons Each black pushbutton corresponds to the prompts on the bottom left and bottom right of the display These buttons are used to access the Settings and Memory Recall menus and scroll up and down between menu...

Страница 4: ...essed Q Erase all memory Erases all stored measurement transactions saved in the meter s memory R S Figure 7 Settings menu R Memory Slot Number Indicates the memory slot number S Resistance Measurement Indicates the resistance measurement value for the respective memory slot Operation General Guidelines Use both 2 27 kg electrodes for Resistance Point to Point Rtt measurements Use one 2 27 kg elec...

Страница 5: ...rounded point 4 Connect the red test lead to one 2 27 kg electrode and place the electrode on the furthest convenient point on the surface 5 Push the red pushbutton to perform a measurement Should the 15 second electrification period appear it may be bypassed by pushing the red pushbutton a second time 6 Push the right black pushbutton to save the measurement and apply an included measurement loca...

Страница 6: ...a once a month Cable jackets should also be cleaned in this fashion Cleaning the 2 27 kg Electrodes Clean the electrodes with a minimum 70 isopropanol water solution Make sure the 2 27 kg electrodes conductive pads are dry prior to use Figure 13 Measuring the Resistance Point to Point Rtt of floor materials Measure Resistance Point to Point Rtt on the Surface 1 Do not clean the surface 2 Remove fr...

Страница 7: ... area needs to be 75 F 3 F 23 9 C 1 7 C at 40 to 60 relative humidity The meter needs to stay at a temperature of 75 F 3 F 23 9 C 1 7 C for approximately 1 hour for proper readings The meter cannot be normalised inside objects enclosed boxes containers or cases The temperature inside an enclosed case will differ from the outside temperature These cases will act as insulators The meter must remain ...

Страница 8: ...cy 10 Relative Humidity Accuracy 10 integers Power Supply 4 AA alkaline batteries Battery Life Approximately 1 500 measurements Display OLED 2 7 diagonal 128 x 64 pixel resolution Memory Capacity 100 measurements Operating Temperature 41 ºF to 85 ºF 5 ºC to 30 ºC Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C Dimension...

Отзывы: