TB-3072

 Page 4 of 4

© 2017 DESCO INDUSTRIES, INC.

Employee Owned

Specifications

Input Voltage and 

Frequency

(External Adapter)

AC/DC Power Adapter
Power Input: 

100-240VAC, 50/60 Hz
Power Output: 

24VDC @ 150mA
Cable Length: 

5 ft. (1.5 m)

Operating Temperature 50 to 95°F (10 to 35°C)
Environmental 

Requirements

Indoor use only at altitudes 

less than 6500 ft. (2 km)
Maximum relative humidity 

of 80% up to 85°F (30°C) 

decreasing linearly to 50% @ 

85°F (30°C)

Dimensions

2.77" L x 2.08" W x .90" H

(70 mm x 53 mm x 23 mm)

Weight

0.1 lbs. (0.05 kg)

Operator Test Range* 500 kilohms to 10 megohms 

impedance

Worksurface Test 

Limit**

500 megohms (±20%)

Operator Test Voltage 1.2V peak-to-peak @ 1.2µA, 

open circuit

Worksurface Test 

Voltage

5 to 7.5V, open circuit

Alarm Response Time <50 ms
Country of Origin

United States of America

*This cannot be verified with standard DC test

equipment.  The continuous monitor is an impedance

sensing device, and the limits are determined by the

magnitude and angle of the impedance.
**The worksurface mat must have a conductive 

layer such as Dual Layer Rubber or Dissipative 

3-Layer Vinyl with conductive buried layers.  Desco

workstation monitors are not recommended for use with

homogeneous matting.

Calibration

Frequency of recalibration should be based on the 

critical nature of those ESD sensitive items handled 

and the risk of failure for the ESD protective equipment 

and materials.  In general, Desco recommends that 

calibration be performed annually.
Use the Desco 

98220

 Calibration Unit for Wave 

Distortion Monitors to perform periodic verification (once 

every 6-12 months) of the Mini Monitor.  The Calibration 

Unit can be used to check the test limits of the Mini 

Monitor without removing it from the workstation.
See 

TB-2082

 for more information.

Figure 6.  Desco 98220 Calibration Unit for Wave 

Distortion Monitors

Содержание 19239

Страница 1: ... a sine wave that the monitor circuit compares to established patterns By monitoring the distortions or shape of the sine wave Wave Distortion Technology determines if the monitored circuit is complete the operator is in the circuit and the total equivalent DC resistance is within specifications Wave Distortion Technology produces a very fast alarm time 50 milliseconds and minimal false alarms ANS...

Страница 2: ...ord from the wristband and attach it to the park snap or remove the banana plug from the jack Upon returning to the workstation and disconnecting the wrist cord from the park snap the operator has 6 seconds to reconnect the wrist cord to a worn wristband before activating the alarm E Monitored Operator Jack Insert the single wire wrist cord here to monitor the operator s connection to ground F 10m...

Страница 3: ...s using the 8 32 button cap screws 5 Align the two snaps underneath the Mini Monitor to the two snaps on the Micastat laminate Apply downward force on the monitor once aligned to snap the Mini Monitor into place 6 Connect the power adapter to the power jack located on the top side of the Mini Monitor Route the wire from the supply to a nearby AC outlet and plug it into the outlet Make sure the vol...

Страница 4: ...erified with standard DC test equipment The continuous monitor is an impedance sensing device and the limits are determined by the magnitude and angle of the impedance The worksurface mat must have a conductive layer such as Dual Layer Rubber or Dissipative 3 Layer Vinyl with conductive buried layers Desco workstation monitors are not recommended for use with homogeneous matting Calibration Freque...

Отзывы: