background image

TB-2039

Page 4 of 4

DESCO WEST

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449  

DESCO EAST

- 90 Hudson Road, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: http://www.desco.com

Limited Warranty

Desco expressly warrants that for a period of one (1) year from
the date of purchase, Desco Calibration Units will be free of
defects in material (parts) and workmanship (labor). Within the
warranty period, the product will be tested, repaired, or replaced
at Desco’s option, free of charge. Call our Customer Service
Department at 909-627-8178 (Chino, CA) or 781- 821-8370
(Canton, MA) for a Return Material Authorization (RMA) and
proper shipping instructions and address. Include a copy of your
original packing slip, invoice, or other proof of purchase date.
Any unit under warranty should be shipped prepaid to the Desco
factory. Warranty repairs will take approximately two weeks.

If your unit is out of warranty, call Customer Service at 909-627-
8178 (Chino, CA) or 781-821-8370 (Canton, MA) for a Return
Material Authorization (RMA) and proper shipping instructions
and address. Desco  will quote repair charges necessary to
bring your unit up to factory standards. 

Warranty Exclusions

THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU
OF ALL OTHER PRODUCT WARRANTIES, EXPRESSED AND
IMPLIED, INCLUDING MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY
DISCLAIMED.  The express warranty will not apply to defects or
damage due to accidents, neglect, misuse, alterations, operator
error, or failure to properly maintain, clean or repair products.

Limit of Liability

In no event will Desco or any seller be responsible or liable for
any injury, loss or damage, direct or consequential, arising out of
the use of or the inability to use the product.  Before using,
users shall determine the suitability of the product for their
intended use, and users assume all risk and liability whatsoever
in connection therewith.

Resistance Value 

Test Output - LED

Red jack “A”  -  675K:

Red (Low-hazard)

Green jack “B”  -  825K:

Green (OK-in limits)

Green jack “C”  -   80M:

Green (OK-in limits)

Yellow jack “D” -  120M:

Red (High resistance)

Step 3:  

Repeat the procedure testing across jacks B,

C (80M), and D (120M).  The different resistance
ranges should give the display shown above.  If the
tester does not meet calibration specifications, contact
factory for calibration, or ask for Technical Bulletin 
TB-2040.

Figure 9. Testing 19250 for “PASS” and “FAIL” ranges
for footwear circuit as indicated on calibration unit.

Specifications

Power Source:

Passive device, no power source
required 

Resistance Ranges -

Wrist straps:

Low - 

675 Kilohm

Pass - 

825 Kilohm

Pass -

8.5 Megohm

High -

11.5 Megohm

Footwear:

Low - 

675 Kilohm

Pass - 

825 Kilohm

Pass -

80 Megohm

High -

120 Megohm

Calibration:

NIST traceable

Calibration Interval:

Recommended annually

Accuracy:

±2%

Weight:

10 ounces

Dimensions:

4.50" x 3.25" x 1.00"

Consider adding the 07010 to your list of “test
equipment that shall be selected to make
measurements of appropriate properties of the
technical requirements that are incorporated into the
ESD program plan” as required by paragraph 6.1.3.1 of
ANSI/ESD S20.20.

© 2003 DESCO INDUSTRIES INC.

Employee Owned

Содержание 07010

Страница 1: ...hat testers are in calibration that are used to verify the parameters of wrist straps and foot grounders The 07010 Calibration Unit easily pays for itself Instead of returning to Desco a Calibration House or even your Metrology Department testers can be NIST calibrated where located in the factory being out of service only a couple minutes No need to have an extra replacement Tester Remove the cal...

Страница 2: ...010 to 98280 Step 2 Connect the second test lead to the 675K jack labeled A on the 07010 Touch the opposite end of test lead to touch tester plate and press down on tester with enough pressure to activate the test circuit Observe the LED s for the proper response as indicated below Be sure to hold the cord at an insulated point so that the resistance value is not affected by the body Resistance Va...

Страница 3: ...cuit needs to be tested individually For detailed information on these testers please ask for Technical Bulletin TB 2040 Wrist Strap Test Circuit Step 1 Connect one of the test leads to common point banana jack labeled E on 07010 Connect the opposite end of test lead to the 19250 s wrist cord jack located on the face of the unit Step 2 Connect the second test lead to the 675K jack labeled A on 070...

Страница 4: ...t misuse alterations operator error or failure to properly maintain clean or repair products Limit of Liability In no event will Desco or any seller be responsible or liable for any injury loss or damage direct or consequential arising out of the use of or the inability to use the product Before using users shall determine the suitability of the product for their intended use and users assume all ...

Отзывы: