TB-6502

  Page 6 of 8

 

                                                     

Revision May 2007

  

 

                                                      

© 2007 DESCO INDUSTRIES INC.                                          

Employee Owned

 

plugged into the phono jack on the front side of the tester.  (See Figure 3.) The 50407/50413 
automatically determines what type of wrist strap is being tested. 

 

The LED(s) will turn off while the test is in progress; the test results for each foot and wrist 
strap will then be displayed for approximately three (3) seconds.  If all test results are “SAFE” 
the internal relay will be activated while the test results are displayed.  The relay can be used 
for opening an electric lock to an ESD sensitive area.  The maximum contact rating is: 
1A@30VDC. 

 

If any of the test results are failed “HIGH” or “LOW” an audible alarm will sound.  The LED(s) 
indicating the failed test will be displayed for approximately three (3) seconds, and the 
internal relay will not activate. 
 

 

CALIBRATION AND PERIODIC TESTING: 

 
The accuracy of the 50407/50413

 

is specified as: 

 
1. +5% for 1 Megohms and lower resistance ranges. 
2. +10% for 1 Megohms and higher resistance ranges. 

 
A periodic check (once every 6 to 12 months) using a precision resistance box should be 
performed to verify proper operation. 

 

A limit comparator, the 50401, can be purchased for the convenient periodic testing of the 
50407/50413. 

 

The 50401 contains 14 standard resistors to check the following high and low limit ranges: 

1. 10K low 
2. 10K safe 
3. 100K low 
4. 100K safe 
5. 1M low 
6. 1M safe 
7. 10M safe 
8. 10M high 
9. 35M safe 
10. 35M  high 
11. 100M  safe 
12. 100M  high 
13. 1G  safe 
14. 1G  high 

 

TESTING PROCEDURE: 
 
Wrist strap operation test. 

 

1.  Insert the 50401/50407/50413 test plug into the dual wrist strap connector on the 

50407/50413. 

Содержание 50407

Страница 1: ...y 2007 2007 DESCO INDUSTRIES INC Employee Owned 50407 50413 DUAL INDEPENDENT FOOTWEAR AND WRIST STRAP TESTER INSTALLATION AND OPERATING INSTRUCTIONS 3651 Walnut Avenue Chino CA 91710 Phone 909 627 8178 Fax 909 627 7449 DescoEMIT com ...

Страница 2: ...g of the wrist strap includes the resistance of the groundable point on the end of the cord the cord itself the resistor the cord to cuff snap connector the resistance of the interface of the cuff the cuff wrist interface and the resistance of the person between the wrist and the hand that contacts the test electrode Per ESD Handbook TR 20 20 paragraph 5 3 2 2 2 Wrist Strap Ground Cord At first gl...

Страница 3: ... High Limit Resistance ON ON 10 Megohms 1 X 10E7 OFF OFF 35 Megohms 3 5 X 10E7 DEFAULT EUROPE ON OFF 100 Megohms 1 X 10E8 DEFAULT USA OFF ON 1 Gigohm At 1 Gigohms high limit resistance a dirty footplate could result in a false pass When using this setting special care must be taken to keep the footplate clean DIP switches 3 4 control the LOW limit for the footwear test selection as follows Switch ...

Страница 4: ...ght Jack is located under foot plate at upper center of foot plate Figure 1 Bottom View Insert the banana plug end of the R391 Banana Plug to Ring Tongue into the GROUND jack on the bottom of the tester See Figure 1 Connect the other end of the cable to earth ground Verify that both connections are tight Failure to correctly ground the tester could significantly degrade the accuracy of the measure...

Страница 5: ...rts the test During the test all LEDs will be OFF to indicate that a test is in progress The touch plate must remain depressed until the test results are displayed Depending on the configuration of the tester the test could require up to three 3 seconds The resistance is checked from the touch plate to the corresponding foot plate for each foot and from the touch plate to the wrist strap connector...

Страница 6: ...icating the failed test will be displayed for approximately three 3 seconds and the internal relay will not activate CALIBRATION AND PERIODIC TESTING The accuracy of the 50407 50413 is specified as 1 5 for 1 Megohms and lower resistance ranges 2 10 for 1 Megohms and higher resistance ranges A periodic check once every 6 to 12 months using a precision resistance box should be performed to verify pr...

Страница 7: ...propriate on the 50401 50407 50413 and repeat the test The results of the test should be fail high for the wrist strap Footwear operation test 1 Insert the 50401 50407 50413 test plug into the foot plate cable connector on the 50407 50413 2 Select the appropriate fail low range on the rotary switch of the 50401 50407 50413 3 Press and hold the touch plate of the 50407 50413 until the test is compl...

Страница 8: ...e include a copy of your original packing slip invoice or other proof of date of purchase Any unit under warranty should be shipped prepaid to the Desco EMIT factory Warranty replacements will take approximately two weeks If your unit is out of warranty Desco EMIT will quote repair charges necessary to bring your unit up to factory standards Call Customer Service at 909 627 8178 for proper shippin...

Отзывы: