background image

English

Instructions

CAUTION: Prior to using the DermLite DL3, read these instructions and 

keep them in a safe place for future reference.

CAUTION: United States Federal Law restricts this device to sale by or 

on the order of a physician.

CAUTION: The DermLite DL3 is designed for external examination only.

DermLite DL3 is a camera-compatible pocket dermatoscope designed to 

view skin lesions with high magnification and clarity.

Two sets of white LEDs (light emitting diodes) can be activated indepen-

dently; one set of LEDs provides cross-polarized light (by canceling the 

reflected light from the skin), whereas the other set of LEDs provides 

non-polarized light for traditional immersion fluid dermoscopy.  

A high quality, four-element, 25 mm, 10x lens with superb color cor-

rection and reduced image distortion produces an image rich in surface 

detail.

To facilitate the use of immersion fluids, and allow for fixed digital imag-

ing, the unit is equipped with a retractable faceplate spacer.

Directions for Use

CAUTION: Do not look directly into the bright LED light. Patients must 

close their eyes during facial examinations.

IMPORTANT: Before initial use, please charge the internal lithium ion 

battery for a minimum of 4 hours using the charging base provided. See 

“Charging” for details.

Hold your DermLite DL3 so that the LEDs are aimed in the direction of 

the lesion to be examined.  Place the device approximately 1” (~25mm) 

from the skin. Push and hold the power button (A) at the right side of the 

unit for approximately 1 second to turn on the cross-polarized mode. A 

quick tap of the button enables the non-polarized mode for immer-

sion fluid dermoscopy, another tap toggles back to the cross-polarized 

illumination.  

Look through the lens with your eye placed about 1 to 4” (25-100 mm) 

above your DermLite. Move the device closer or further away from the 

skin to obtain the desired image focus. For dermoscopy with skin con-

tact, or for camera use, rotate the Focus Dial (B) to the left or right. The 

position “0” indicates the ideal distance for photography. To turn off your 

DermLite DL3, push & hold the power button for approx. 1 sec.

CAUTION: If your DermLite is used for oil immersion imaging, place the 

oil only on the skin and then place the extended faceplate spacer on top 

of the oil.  Do not allow oil to contact the charging contacts or battery 

compartment.

Removing the Spacer

Unscrew the Retaining Ring (B1) and rotate the Focus Dial (B) until the 

Spacer (B2) drops. 

To re-attach the Spacer, slip its three pins through the three notches (D) 

and gently rotate the Focus Dial until the Spacer is being pulled back in. 

Reattach the Retaining Ring.

Eyepiece

To reduce internal reflections and improve the contrast of the image, 

attach the optional eyepiece onto the 28mm thread (E).

Camera Use

To enable you to attach your DermLite DL3 to a camera, the device is 

equipped with a 28 mm thread (E). A simple ring adapter (such as a 

28-37 mm stepping ring) may be all that is required to adapt the device 

to your camera of choice, although many cameras require a special 

adapter made by the camera manufacturer or a photography supply 

company. Prior to taking images, wipe the Faceplate (F) clean and extend 

the Spacer to the “0” position. Generally, good images may be obtained 

by setting your camera to automatic or “P” mode, activating the “macro” 

function, and deactivating the flash. Please contact 3Gen or your camera 

dealer for more information.

Charging

Prior to using the DermLite, insert the 110V-220V power plug into an 

electrical outlet and the other end into the back of the charging base. 

Set the base on a level surface. Place the DermLite onto the base, and 

the “3” will illuminate red to indicate that the DermLite DL3 is being 

charged. When the lithium battery is fully charged and either in its base 

or outside its base and powered on, the “3” is green and the device will 

allow up to 2 hours of continuous operation, or enough power for imag-

ing approximately 600 lesions at 30 seconds per examination. If less than 

25% battery life remains, the “3” will turn orange.

 

After approximately 1,000 charges, the battery’s capacity will diminish 

and the battery may need to be replaced. To remove the battery, slide 

the rubber handle (G) back and completely off the device. Disconnect the 

small white battery plug and remove the battery. Insert a new battery 

and reconnect the battery plug. Slide the rubber handle securely onto 

the device. 

CAUTION: Your DermLite DL3 uses a special lithium battery, which can 

only be purchased from 3Gen or from an authorized 3Gen dealer. Do 

not under any circumstances use another battery other than the one 

designed for this unit. Doing so could result in damage to the unit. Also, 

do not under any circumstances use another charger other than the one 

supplied with the unit.

Using the Lanyard

CAUTION: Lanyards pose a strangulation risk. Use with caution.

Run the short, thin portion string through the eyelet (H) located at the 

end of the device. Feed the other end of the lanyard through the loop 

made by the thin string (I). Separate the DermLite from the lanyard by 

squeezing the snap connector (J).

Troubleshooting

Please check our website (www.dermlite.com) for the most current trou-

bleshooting information. If your DermLite requires servicing, please go 

to the “Service & Support” pages at www.dermlite.com, where you can 

complete a “Repair Request” form. Or, contact your local 3Gen dealer.

No Light

Fully recharge the battery and check again. If the condition persists, your 

DermLite may require servicing.

LED Fails

The LEDs used in your DermLite are designed to last over 100,000 hours. 

If any one of the LEDs fails, your DermLite requires servicing.

Mechanical Malfunction

Please contact 3Gen to arrange for service.

Care and Maintenance

Your DermLite DL3 is designed for trouble-free operation. There are 

no adjustments to be made nor is any special servicing required. Never 

attempt to open the device for any reason.

Cleaning and Sterilization

The exterior of your DermLite may be cleaned or disinfected by wiping it 

with Isopropyl alcohol (70% vol.). Do not use alcohol or disinfectants in 

the optical areas of the unit. Do not use abrasive material on any part of 

the equipment or immerse the device in liquid. Do not autoclave.

Cleaning the Lens and the Polarizers

The lens and the polarizing filters should be treated as high quality pho-

tographic equipment and should be cleaned with standard lens cleaning 

equipment and protected from harmful chemicals.

 

DermLite DL3 components include:

Twenty-eight 3mm LEDs, four-element 25mm lens with 10x magni-

fication, cross-polarized and non-polarized illumination, retractable 

faceplate spacer with 10 mm scale, 28 mm eyepiece, lanyard

Warranty

5 years for parts and labor.

Serial number and company information are located inside the battery 

compartment.