Denver Instrument TB-124 Скачать руководство пользователя страница 23

Data ID Codes

To help you identify weights, piece counts, percentages, etc., a code letter
is printed or displayed in front of these values. For example, an “N”
printed or displayed in front of a weight value identifies it as a net weight.
If you set the code for “without data ID code,” only net weights, results in
percent and counting results will be output. You will find the data ID codes
for a particular application program listed in the corresponding
description.
The ID code increases the data output format for each weight readout from
16 to 22 characters.

- without ID code 

(Menu Code 7 2 1*)

- with ID code 

(Menu Code 7 2 2)

ISO/GLP-compliant Printout or Record

Use of the balance as a test and measuring instrument in quality
assurance systems in compliance with the requirements of ISO, GLP, GMP
and EN (European Standards) in which proof of the balance’s accurate
performance is required. The balance can record all completed calibration
or adjustment operations and print out data in compliance with the
requirements of Good Laboratory Practice (GLP). The balance, interfaced
with a data printer or a computer, can create a document that records the
date, time, serial number and model number, making it possible to clearly
trace data to the balance that generated it and the time at which it was
generated.

Select the ISO/GLP-compliant printout or record mode by setting the
respective code in the balance operating menu:
ISO/GLP-compliant printout/record mode 

- Off 

(Menu Code 8 10 1*)

- only for adjustment/calibration and linearization functions 

(Menu Code 8 10 2)

- always on (e.g., for adjustment/calibration and linearization functions,

application programs, weight readouts) 

(Menu Code 8 10 3)

The following menu code setting must be selected in order to obtain an
ISO/GLP- compliant printout/record:

- with data ID code 

(Menu Code 7 2 2)

- with application parameter

(Menu Code 7 1 2)

ISO/GLP-compliant printouts/records will not be generated if the
factory setting, menu code 7 2 1 or 711  is selected. In addition, do
not select the “Auto print” data output parameter
(Menu Code  6 1 4 or 6 1 5).

19

Memo:

Note:

Содержание TB-124

Страница 1: ...TB Series Operation Manual 0 000 g FUN CTIO N ON OFF CAL CF PRIN T TARE SI 413 902535 1 Rev A 0 0 g FUN CTIO N ON OFF CAL CF PRIN T TARE SI 6201 0 0000 g FUN CTIO N ON OFF CAL CF PRIN T TARE SI 124...

Страница 2: ...An instrument can be no more accurate than the standard to which it has been compared For assistance in the selection of reference weights please contact the factory Caution Changes or modifications n...

Страница 3: ...Calibration 6 Block Calibration 7 Percent Weighing 7 Counting 9 Animal Weighing 10 Tare Memory 12 Balance Operating Parameters 12 Ambient Conditions 12 Stability Signal 13 Tare Parameters 13 Auto Zero...

Страница 4: ...Dimensions LxWxH Toploading 11 8 x 7 8 x 3 5 30 x 20 x 9cm Net Weight Analytical 13 2 lbs 6 0 kg Net Weight Toploading 10 0 lbs 4 5 kg Power Requirements 15 VDC 100 mA with AC adapter center pin Tempe...

Страница 5: ...Before packing your balance unplug all connected cables to prevent damage Do not expose the balance unnecessarily to extreme temperatures moisture shocks blows or vibration Remove the plastic wrapping...

Страница 6: ...e components listed below on the balance in the order given Pan support Weighing pan Glass draft shield cylinder Draft shield cover Preparing Balances with a Square Weighing Pan Place the components l...

Страница 7: ...l safety precautions The ground terminal is connected to the balance housing which can be additionally grounded for operation The data interface is also electrically connected to the balance housing g...

Страница 8: ...lt setting when you press the ON OFF key to turn off the display the balance will remain in the standby mode This means that it will be ready to operate without requiring any warm up when you turn the...

Страница 9: ...oz 0 035273962 1 7 6 3 1 6 Troy ounces ozt 0 032150747 1 7 7 3 1 7 Hong Kong taels tl thl 0 02671725 1 7 8 3 1 8 Singapore taels tl ths 0 02646063 1 7 9 3 1 9 Taiwanese taels tl tht 0 02666666 1 7 10...

Страница 10: ...3 When display reads 0 grams press the CAL key to activate the calibration function The built in calibration weight is internally applied by servomotor and removed at the end of adjustment or calibrat...

Страница 11: ...balance 4 Press the FUNCTION key and the current reference percentage will appear on the screen 5 The screen will then flash and show the percentage on the balance 6 Add or subtract parts and the per...

Страница 12: ...rr 22 may briefly be displayed This means that the new reference value has been stored Storage Parameter for the Reference Weight Value The reference weight value is stored with full accuracy accordin...

Страница 13: ...esetting the Reference Weight 1 Place the new reference weight on the balance 2 Hold down CF for more than 2 seconds until the reading is in grams 3 Press FUNCTION and rEF current setting pcs appears...

Страница 14: ...symbol AUTO will be displayed 3 A count down begins for the number of readings used for averaging 4 When the result is locked the symbol and AUTO flashes 5 The icon will stop flashing after you have u...

Страница 15: ...ights are within the previously selected range If so the actual averaging operation will be started Delay start until difference is slight Menu Code 3 7 1 difference is average Menu Code 3 7 2 differe...

Страница 16: ...N gross weight tare net weight Balance Operating Parameters In the operating menu of the balance you can define how your balance will adapt to ambient conditions and also how it will work to meet you...

Страница 17: ...1 not until the readout is stable Menu Code 1 5 2 Auto Zero Function When this zero tracking function is activated any slight changes off the zero readout are automatically tared auto Zero On Menu Cod...

Страница 18: ...Weighing Capacity of 10 kg To open the below balance port remove the cover plate from the bottom of the balance Now you can attach a sample using a suspension wire for example Common applications for...

Страница 19: ...or output data press the PRINT key Interface Description This description has been written for users who wish to connect their balance which has a built in V24 V28 RS 232C S 423 interface port as a s...

Страница 20: ...ecting the cable and disconnect any lines marked Internally Connected e g pin 6 Failure to do so may damage or even completely ruin your balance and or peripheral device Pin Assignment Pin 1 Signal Gr...

Страница 21: ...rt on the balance this will not generate an error code In this case data will be output but not received Data Bits 7 not adjustable Baud Rate Code 150 baud 5 1 1 300 baud 5 1 2 600 baud 5 1 3 1 200 ba...

Страница 22: ...for 2 display updates must be set if you connect the remote display to the balance 1 display update Menu Code 6 3 1 2 display updates Menu Code 6 3 2 Automatic Taring after Data Output This convenien...

Страница 23: ...ustment operations and print out data in compliance with the requirements of Good Laboratory Practice GLP The balance interfaced with a data printer or a computer can create a document that records th...

Страница 24: ...ed in the data string Control Commands with Upper Case Letters or Special Characters ESC P CR LF Print activate block auto print ESC S CR LF Restart self test ESC T CR LF Tare ESC P CR LF Internal cal...

Страница 25: ...o be activated by commands ESC f 0 _ CR LF function key ESC f 1 _ CR LF cal key ESC s 3 _ CR LF clear function ESC x 0 _ CR LF Perform calibration test ESC x 1 _ CR LF Output scale model ESC x 2 _ CR...

Страница 26: ...Position the pan shows L is not in place or Err 54 The weight Too much vibration Access the menu readout changes or the balance to select the correct constantly is exposed to a draft code for the wei...

Страница 27: ...mediately Lock the equipment in a secure place to ensure that it cannot be used for the time being Safe operation of the balance with the AC adapter is no longer ensured when there is visible damage t...

Страница 28: ...ment promptly upon your purchase of the Denver Instrument product The return of the card is not a condition precedent to warranty coverage 2 If you have any questions about a Denver Instrument Summit...

Страница 29: ...e number on display to desired value until 5 appears 6 Press PRINT to move to the 3rd number when you move to the third number the previously set menu code will appear 7 Press CAL CF to increase numbe...

Страница 30: ...423 4831 U K and Ireland Denver Instrument Company Denver House Sovereign Way Trafalgar Business Park Downham Market Norfolk PE38 9SW England Tel 44 136 63862 42 Fax 44 136 63862 04 Europe Asia and Au...

Отзывы: