Troubleshooting
Display Shows:
Cause:
Remedy:
No segments appear
I/O key pressed
Press the I/O key.
on the display.
to turn OFF Display.
L
L
The balance is
Check the weighing pan
underloaded.
for proper installation.
Weight is too light or there
Increase reference
is no sample on balance.
quantity.
H
H
The load exceeds the
Unload the balance or
balance capacity.
look for obstruction.
Display capacity
Decrease weight
is exceeded.
on balance.
L
LO
O
Low battery.
Replace battery.
E
E --0
02
2
The calibration operation
Place balance in a
is being attempted and
stable environment
vibration, air drafts or
before calibrating.
other unstable conditions
Before pushing the
exist.
CAL/MODE button, the
LCD should display a stable
ZERO with the arrow mode.
E
E --110
0
There is a value stored
Clear the memory or
in the memory when the
turn the balance off to
TARE button is pressed.
restore normal tare function.
E
E --3
30
0
An electronic malfunction
Contact Denver Instrument
has occurred.
E
E--5
54
4
This indicates the
Contact Denver Instrument
electronics are no longer
within the factory set
parameters. Objects have
dropped onto the weighing
pan or the balance itself
has been dropped.
▲
The sensor is out of
Contact Denver Instrument
specification. The balance
has been damaged by
rough handling. Liquids or
other materials have spilled
into the balance.
Do not attempt to access internal components or make any unau-