7.3.4
Arm opens and does not hold capsule.
Possible Cause
Solution
A Broken arm
Replace arm assembly (Fig 6c). See
assembly/disassembly.
7.3.5
Rubber boot falls off.
Possible Cause
Solution
A. Glue bond is broken
Glue the boot back to plastic using Loctite Super
bonder 498 (Fig 8.2).
7.3.6
Window makes noise during operation.
Possible Cause
Solution
A. Tension of the window has decreased. Locate the tension springs and switch plate
(Fig 6d). Change the tension springs, switch plate
and the window cover.
15
Figure 6d
Figure 8.1
Tension Spring
Tension Spring
Switch Plate
Middle Screw