Technical specifications
3.1 General
Parameter
Value
Unit
External dimensions (W × H × D)
385 × 780 × 500
mm
Heating chamber volume
2 bowls
120 mm × 35 mm
Maximum temperature
1650 °C
Heating rate
selectable: 1
–99 °C/min
Weight
58 kg
3.2 Electrical connection
Electrical
Parameter
Value Unit
Nominal supply voltage
220-240 V AC
Frequency
50 Hz
Permissible deviation
from the nominal voltage
± 10 %
Maximum Power
consumption
2000 W
Overvoltage category
II (pursuant to IEC 60364-4-443)
Protection against
overvoltage
Provide suitable surge filters
Fuse rating (on site)
Connect to a dedicated circuit protected by a
T10 AH 250 V fuse
3.3 Operating conditions
Environment
Parameter
Value Unit
Temperature range
5
–40 °C
Maximum relative
humidity
up to 31°C: 80 %
up to 40°C: 50 %
non-condensing
Installation
Indoors in dry rooms only
Height above sea level
maximum 2000 m
Pollution severity
Tested for pollution severity
– Pollution degree II