CAUTION: Never operate the handpiece under the
following conditions:
1. In the presence of a soft tissue wound.
2. With a bur extended out from handpiece head.
3. Under "no load" conditions for extended periods of time.
4. At air pressure exceeding 40 p.s.i. (2.7 Atm.) for 180S,
360S, 540S series and 42 psi (2.9 Atm.) for 181S, 361S,
541S series.
5. With a loose or bent bur.
6. If you experience a sudden and noticeable change in
handpiece noise or vibration.
7. If bur moves forward more than 1 mm while
operating handpiece.
Recommended Burs & Diamonds For:
180S and 360S Series
• Regular Friction Grip (FG) burs
• Shaft diameter of 1.6 mm per ISO 1797-1
• Total length not to exceed 23 mm
540S Series
• Regular Friction Grip (FG) or Short Shank (SS) burs
• Shaft diameter of 1.6 mm per ISO 1797-1
• Total length not to exceed 23 mm
9
Operation,
continued