Insert handpiece (without a bur) into sterilization bag and load into
autoclave. (DENTSPLY Professional recommends using sterilization
bags with at least one paper side.) Handpieces should be processed
in the autoclave for 15 minutes at a sterilization temperature of
135ºC/275ºF
.
For cassette chamber saturated steam devices, handpieces
should be processed unbagged for 3.5 minutes at a
sterilization temperature of
135ºC/275ºF
.
Keep handpiece in bag until ready to use in the operatory.
Insert the coupler into the handpiece until it “clicks” into place.
Check for secure connection by tugging the handpiece. To operate
the handpiece, simply depress the foot control on your dental unit.
Note:
The Midwest Stylus uses delivery unit chip air to deliver fine
mist water spray. If you get a stream of water instead of a fine mist,
just increase the “chip air” or “coolant air” on the delivery unit until
the desired fine mist spray is achieved (see page 5 for recommended
settings).
Operation
8
Bag and Autoclave