background image

SmartLite™ PS

  

Scientific 

Compendium 

 

Barcol hardness: Spectrum TPH A2 (10 sec exposure)

50

55

60

65

70

75

80

934

-h

ar

n

d

es

s

80

85

90

95

100

105

b

o

tt

om

/t

op

 [

%

]

top hardness

72

72

72

73

72

bottom hardness

70

71

69

69

71

bottom/top [%]

97

99

96

95

98

SmartLite PS

Epilar FreeLight 2

Starlight Pro

L.E.Demetron I

Spectrum 800@800

 

Figure 8 

Barcol hardness of Spectrum TPH A2 after a 10sec cure with various curing 
units 

Barcol hardness: Spectrum TPH A4 (10 sec exposure)

50

55

60

65

70

75

80

934-

h

ar

n

d

ess

80

85

90

95

100

105

bo

tt

om

/t

o

p

 [

%

]

top hardness

72

72

73

73

70

bottom hardness

66

70

64

67

66

bottom/top [%]

92

97

88

92

94

SmartLite PS

Epilar FreeLight2

Starlight Pro

L.E.Demetron I

Spectrum 800@800

 

Figure 9 

Barcol hardness of Spectrum TPH A4 after a 10sec cure with various curing 
units 

6.3 Compressive 

strength 

Additionally, measurements of the curing performance of the SmartLite PS were 

compared to that of the Spectrum800@800 by measuring the compressive strength of 

EsthetX A2 and A4 and Spectrum TPH A2 and A4 shades, respectively. Results are 

shown in Figure 10. The obtained values for the compressive strength of the cured 

materials confirm that there is no difference between the curing performance of the 

SmartLite PS and the Spectrum800 at 800 mW/cm

2

.  

 

 

11 / 22 

Содержание DeTrey SmartLite PS

Страница 1: ......

Страница 2: ...erformance 9 6 1 Depth of Cure according to ISO 4049 2000 9 6 2 Barcol hardness 10 6 3 Compressive strength 11 7 Clinical data 12 7 1 Consumer preference evaluation 12 7 2 Clinical post marketing surv...

Страница 3: ...time Thus this technology is an ideal basis for cordless devices improving the convenience of the curing procedure in the dental surgery Noisy and uncomfortable cooling fans necessary in traditional...

Страница 4: ...at 10 seconds with a stable power output even at a low battery charging status This is achieved by means of a microprocessor controlled technology 2 3 Charger Four charger types are available adjusted...

Страница 5: ...n QTH lamp emit light with wavelengths beyond the absorbance range of camphorchinone CQ CQ is the most commonly used photo initiator in almost all visible light cured dental materials Thus the light e...

Страница 6: ...ours and therefore does not require maintenance or constant monitoring As shown in Figure 3 a fully charged battery provides approx 150 ten second cures approx 25 minutes operating time with constant...

Страница 7: ...nducted at the Department of Oral Rehabilitation Medical College of Georgia head Professor F Rueggeberg DDS MS The purpose of this investigation was to simulate and determine in vitro the intra pulpal...

Страница 8: ...ve shown that the SmartLite PS emits and transfers less heat onto the tooth than other widely used LED and halogen curing lights 5 2 Temperature of the LED Tip ISO 60601 requires that equipment parts...

Страница 9: ...s and a strong halogen light according to the method described in the ISO 4049 2000 It is important to note that this particular method defines the depth of cure as 50 of the actual thickness of the c...

Страница 10: ...ter 10sec exposure 6 2 Barcol hardness Another method to determine the curing performance of a curing light is to measure the hardness of the exposed surface and the hardness at a given depth using Ba...

Страница 11: ...ness 72 72 73 73 70 bottom hardness 66 70 64 67 66 bottom top 92 97 88 92 94 SmartLite PS Epilar FreeLight2 Starlight Pro L E Demetron I Spectrum 800 800 Figure 9 Barcol hardness of Spectrum TPH A4 af...

Страница 12: ...Spectrum 800 800 7 Clinical data 7 1 Consumer preference evaluation General dental practitioners were given the opportunity to use SmartLite PS for standard restorative treatment over a time period o...

Страница 13: ...unit Very Poor Poor Satis factor y Good Very good Table 1 Consumer Preferences Consumer Preference Evaluation What practitioners liked about SmartLite PS Small size Easy to use Light in weight Well ba...

Страница 14: ...ance of posterior composite restorations is the same when cured with a a conventional halogen curing light with 20 second curing cycles b SmartLite PS with 10 second curing cycles Design Prospective c...

Страница 15: ...should be no hypersensitivity or sensitivity to percussion For inclusion into the study the following additional requirements have to be met Successful cavity isolation no contamination Patient was ex...

Страница 16: ...ing that the SmartLite PS fully equals strong QTH curing lights like the Spectrum 800 with regard to curing performance The convenient size the ease of use and the device s lightness makes it an ideal...

Страница 17: ...ges different to the range indicated on the charging unit Note To disrupt the power supply the charging device has to be plugged out Always unplug the charger before disinfecting Precautions Anyone wi...

Страница 18: ...n of the light to the area being treated 10 Directions for use 10 1 Installation and charging 1 Plug in charging unit 2 Insert the hand piece 1 into the charging unit 3 The battery light will be flash...

Страница 19: ...hand piece is inserted into the charger SmartLite PS is still being re charged Fast flashing red 2 minutes exposure time left SmartLite PS needs re charging Slowly flashing red Battery is fully disch...

Страница 20: ...evention of cross infection cleaning disinfection and storage The use of the sleeve is an additional precautionary measure against contamination and does not substitute disinfection of the device Afte...

Страница 21: ...y of the rechargeable battery discharge the device by regular use at least every 6 months and recharge the battery immediately thereafter High voltages are present inside the charger Do not open charg...

Страница 22: ...midity 30 80 non condensing Atmospheric pressure 500 1060 hPa Transport Ambient temperature 10 60 C 14 140 F Relative humidity 30 80 non condensing Atmospheric pressure 500 1060 hPa Battery performanc...

Отзывы: