29
FLOW - Delivery plunger does not retract:
•
Shut device off, and then turn back on.
•
Press the
button.
•
Press the activation cuff to move the plunger forward then press the
button.
FLOW - Motor stops turning:
•
The cannula may have been placed too firmly in the canal, replace the
cartridge.
•
Reduce the pressure applied to the cannula.
•
Check to make sure the cartridge nut has not been over-tightened.
Frequently Asked Questions
What size is the smallest EHP? What is it made from?
•
The black EHP is a 40/.03, which is the smallest size in the Calamus
®
line.
The EHPs are stainless steel.
How hot can a tip get?
•
A tip can reach 400°C. The heat can be adjusted as low as 100°C, the default
temperature for the Pack handpiece is 200°C.
What is the normal PACK power setting?
•
The maximum heat setting should only be used only when searing off gutta-
percha or an obturator. For warm vertical compaction or any other downpack
technique, we recommend starting at 200°C.
How long will an EHP last?
•
It depends on how well the EHP is care for. The life of an EHP will be reduced by:
ª
Running it for long periods of time.
ª
Working constantly at maximum temperatures.
ª
Bending or putting too much mechanical force on the EHP.
•
The EHPs should be used only to transfer heat – not to exert manual force!
•
TDS recommends that a manual plugger be used when manual force is
necessary.
•
All tips will gradually lose their heating efficiency over time. This is strongly
dependent on the user and will vary.
Can the Calamus
®
Dual be placed in my medical cart?
•
The Calamus
®
Dual is not designed to be taken out of the UL approved
console.
•
Doing so, can alter or even negate the efficacy and safety designs that the
Calamus
®
Dual was manufactured under. Furthermore, removing the
Calamus
®
Dual from its UL approved housing will void the warranty.