DIRECTIONS FOR USE

3D Obturation System

Содержание Clamus DUAL

Страница 1: ...DIRECTIONS FOR USE 3D Obturation System...

Страница 2: ...lling Hills Drive Johnson City TN 37604 USA 1 800 662 1202 Fax 1 800 597 2779 www tulsadentalspecialties com Manufactured By Aseptico Inc 8333 216th St SE Woodinville WA 98072 USA Manufactured under U...

Страница 3: ...Heat Shield 19 Cartridges 19 Electric Heat Pluggers Thermal Response Tip 20 Maintenance 21 CLINICAL TECHNIQUES 22 Cone Fit Plugger Selection 23 Calamus Downpack Vertical Condensation 24 Calamus Downpa...

Страница 4: ...s for use with the Flow handpiece Bending Tool for the cartridges Cleaning Brush for the Flow handpiece Additional Cartridge Nut for the Flow handpiece Heat shield for the Flow handpiece Package of 10...

Страница 5: ...mum continuous heating time 10 seconds for temperatures above 200 C 15 seconds for temperatures at or below 200 C One minute in thermal response mode Caution This device has been tested and found to c...

Страница 6: ...cuments Class II Equipment Dangerous Voltage Alternating Current Hot Surface Nonreturnable if Seal Is Broken Caution this product contains natural rubber latex which may cause allergic reactions Cartr...

Страница 7: ...bjects a tooth to heat in order to determine pulp vitality The Flow handpiece is intended for heating and extruding gutta percha into the root canal system The single use cartridges are designed to pl...

Страница 8: ...e hot during use Turn off unit and allow cooling before exchanging tips Equipment not suitable for use in the presence of a flammable mixture with air oxygen or nitrous oxide Risk of fire Do not conta...

Страница 9: ...se The heat shield included with the system may be used optionally to reduce the handpiece surface temperature If the heat shield is not utilized avoid contact with the forward portion of the Flow han...

Страница 10: ...V For replacement press on the black plastic clip of the fuse compartment pull out and replace with the correct fuse 2 Remove packaging from handpieces Figure 2 Component Console Identification Flow...

Страница 11: ...the power cord to the back of the console and plug into a grounded electrical outlet 7 Heat Shield The heat shield included is designed to reduce the Flow handpiece surface temperature If the heat shi...

Страница 12: ...ocedure the handpiece cartridge nut and the expended cartridge will be hot to the touch 11 Heat Shield Place the heat shield over the cannula and handpiece Then turn the heat shield until the Flow ind...

Страница 13: ...evice Press the handpiece cuff or any button for the Pack or Flow side of the console The blue LED indicator at the top of the device will light indicating which side is active At initial power on the...

Страница 14: ...ip temperature is preset to 90 C and is not adjustable Volume Settings Press the button and then use the buttons to set the desired audible tone volume The LED light to the right of the button will il...

Страница 15: ...ease decrease the flow rate in 10 increments Holding down one arrow button will result in rapid scrolling through the flow range 20 minimum to 100 maximum Flow rate is displayed on the numerical LED w...

Страница 16: ...ings of 200 C or lower and 10 seconds for settings above 200 C before timing out The Thermal Response Tip will heat for one minute while the 360 activation cuff is depressed To restart after a timeout...

Страница 17: ...a against the previously packed gutta percha for 5 seconds Hold the handpiece lightly when expressing material to allow the device to readily back out of the canal As the material is expressed the ind...

Страница 18: ...to cool CAUTION Do not remove a hot cartridge 5 Unscrew and remove the Flow handpiece cartridge nut See Fig 10 6 Using the hole in the included bending tool remove the cartridge from the handpiece 7 D...

Страница 19: ...h 250V 600 mA rated fuses installed for a 115VAC power source If power source is 230 VAC ensure that 250V 300 mA rated fuses are installed Warning Turn the power off and unplug the unit before followi...

Страница 20: ...from the middle of the cable out to the handpiece and then again out to the console Avoid gripping the cable tightly as this will damage the wire casing over time Handpieces Clean the exterior of the...

Страница 21: ...cartridge on a patient wipe the cannula with alcohol or a disinfectant Make sure that the device has been switched off and cooled down Store cartridges at room temperature Do not immerse the cartridge...

Страница 22: ...running water for a minimum of 1 minute to remove all chemicals 5 Dry thoroughly 6 Steam autoclave plugger tips for 15 minutes at 132 C before initial use and between each patient 7 Inspect the EHPs...

Страница 23: ...and press the button on the console to fully retract the plunger 2 Allow the heater section to reach operation temperature 180 C 3 Turn off unit 4 Insert the included cleaning brush into the heating...

Страница 24: ...though the device is able to precisely control the heat transmission to the tip it is unable to prevent undesired heating up of the root It is recommended to practice and test your technique on extrac...

Страница 25: ...ghtaway portion of the canal and to within 4 5 mm of the canal terminus 5 Select the Calamus Electric Heat Plugger EHP that will passively fit through the straightaway portion of the canal and optimal...

Страница 26: ...circumference of the canal using short firm strokes to again move gutta percha apically clean the canal walls and flatten the material 6 Use this medium Calamus manual plugger and press for 5 seconds...

Страница 27: ...conds to prevent thermal injury 5 Deactivate the Calamus EHP and continue to maintain firm apical pressure on the cooling instrument until the silicone stop reaches the reference point 6 Maintain firm...

Страница 28: ...pressure for 5 seconds to three dimensionally compact the warm mass into this region of the root canal system 6 Continue to activate and progressively pack warm gutta percha deeper into the canal unt...

Страница 29: ...n of the canal and to offset shrinkage during the cooling phase 5 Position the tip of the warm cannula against the previously packed filling material for 5 seconds 6 Activate the Flow handpiece and di...

Страница 30: ...w the plunger to recede this will relieve pressure on cartridge nut FLOW Gutta percha does not flow through cannula Make sure the cartridge is new and not from a previous use Verify that the cannula i...

Страница 31: ...used only when searing off gutta percha or an obturator For warm vertical compaction or any other downpack technique we recommend starting at 200 C How long will an EHP last It depends on how well th...

Страница 32: ...s to notify TDS Customer Complaint Department promptly Customer Complaint will provide instructions usually directing that the product be returned for service Shipment to TDS and the cost thereof is a...