Getting Started
1.
Disconnect the charger from the device if connected
2.
Connect the measurement cable and switch the device On – after a short logo
presentation, the main screen is displayed and the cursor on the tooth image will
start blinking.
(While turning the device On without the measurement cable –
the
main screen
will be displayed without cursor on the tooth image)
3.
Attach the lip clip to the patient
4.
Insert the file into the canal – (to ensure precise measurements the file size
should be adjusted to the canal diameter)
5.
Connect the hook to the file
The cursor on the tooth will stop blinking (accompanied by a double beep signal).
A blinking cursor and no sound signal indicates a faulty connection. Check your
connections, clean your hook connection, moisten the canal, if necessary, and start
again.
No other adjustments are necessary before starting measurement.
Note: Connection test feature is included in
ProPex
®
II
in order to check the cables:
●
Connect metal part of the connection hook
to the lip clip
●
“Connection test” icon
should
appear in the bottom left corner of the
display, indicating proper connection
●
If no icon appears, the connection hook or
the measurement cable should be replaced