8 Maintenance
Operating instructions Midwest Stylus Plus / Midwest Tradition
60
67 27 148 D3673.201.09.01.02
8.2.2
Replacing the compression ring
C
In order to guarantee proper functioning of the “anti-retraction valve
in the spray water duct”, we recommend replacing the compression
ring (C) every six months.
1.
Purge the spray water duct at the maximum water flow.
2.
Slide the tool, with the new compression ring attached, to just
in front of the corresponding groove (C).
3.
Insert the compression ring. Do not use any sharp objects.
4.
Check that the ring fits snugly all the way around and is not
twisted.