Operating instructions Midwest Stylus Plus / Midwest Tradition
6 Postprocessing
67 27 148 D3673.201.09.01.02
41
English
5.
Transport the high-speed handpiece and the quick coupling to
the hygiene room in a suitable transport container.
6.
Conduct automatic reprocessing. Manual reprocessing is
possible in exceptional cases if the national/local regulations
are followed.
NOTE! Do not conduct automatic reprocessing of the quick
coupling.
7.
Lubricate the high-speed handpiece with spray [→ 47].
8.
Sterilize the high-speed handpiece and the accessories
[→ 54].
NOTE! Do not sterilize the quick coupling.
6.2
At the end of the work day
➢