
Operating Instructions Midwest Stylus
®
Plus / Midwest
®
Phoenix ™ /Pro
7 Conditioning
6607290 D3673
.
201.01.03.02
47
1. Brush the high-speed handpiece under running water (< 38 °C,
< 100 °F, at least drinking water quality) and good lighting (min. 500
lux) and color rendering index (min. 80 Ra) until no more dirt can be
seen, for at least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and
adapters according to manufacturer's instructions.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds.
5. Conduct thermal disinfection or unwrapped steam sterilization.
6. Maintain mechanical parts manually [ → 48].
7. Maintain the push button chuck manually [ → 51].
8. Pack the high-speed handpiece in packaging material suitable for
sterilization and storage. e.g. paper/plastic composite packaging.
9. Conduct sterilization [ → 54].
6607290.book Seite 47 Mittwoch, 13. September 2017 9:53 09
Содержание Midwest Phoenix
Страница 2: ...6607290_Titel fm Seite 2 Mittwoch 13 September 2017 11 01 11...
Страница 4: ...6607290 book Seite 2 Mittwoch 13 September 2017 9 53 09...
Страница 74: ...6607290 book Seite 72 Mittwoch 13 September 2017 9 53 09...
Страница 146: ...6607290 book Seite 72 Mittwoch 13 September 2017 10 37 10...
Страница 220: ...6607290 book Seite 74 Mittwoch 13 September 2017 10 52 10...
Страница 221: ...09WW Ruck fm Seite 1 Mittwoch 13 September 2017 10 58 10...