8 Maintenance
8.1 Care, cleaning, disinfection, and sterilization
64 32 087 D3492
D3492.201.07.24.02 07.2018
57
8.1.5.7
High-level disinfection of the mirror sleeve
NOTE
The hot air sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection (HLD) is as follows –
assuming the CIDEX
®
OPA disinfectant is available in your country via
your dealer:
A HLD set to support the HLD processes can be ordered from Dentsply
Sirona with REF 66 05 120. This replaces the set with REF 63 46 907.
The processes for both HLD sets can be found in the document
"CEREC camera – care, cleaning, disinfection and sterilization”
available on our online portal for Technical Documentation
(http:// www.dentsplysirona.com/manuals).
A
HLD tank
C
Metal bracket
B
Cover
D
Plugs
1.
Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve [→ 56]").
2.
Always use personal protective equipment when implementing the
high-level disinfection if not before.
3.
Remove the mirror sleeve from the CEREC Omnicam (see
Removing the mirror sleeve [→ 55]).
4.
Place the white protective cap on the camera head and place the
camera in the camera storage unit.
5.
Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.