8.1.5.1
General information
The CEREC Omnicam is a very sensitive optical device and must
therefore be handled with the utmost care. Protect the coated sapphire
glass and the camera window against scratches and clean them with a
lint-free cloth and ethanol (commercially available cleaning alcohol) if
any haze is noted during the acquisition.
The following methods are available for cleaning the mirror sleeves:
● Wipe disinfection for the camera and mirror sleeve [→ 56]
● High-level disinfection of the mirror sleeve [→ 57]
● Hot air sterilization of the mirror sleeve [→ 61]
8.1.5.2
Components of the Omnicam
A
Press detent to release
C
Sapphire glass (coated)
B
Mirror sleeve
D
Camera window
8.1.5.3
Removing the mirror sleeve
If required to remove the mirror sleeve do the following:
1.
Press the mirror sleeve against the camera body.
2.
Press detent A.
NOTE
There is a risk of damaging the camera window or the coated
sapphire glass if the mirror sleeve is not pushed straight toward the
front.
➢ Push the mirror sleeve straight toward the front; do not tilt it.
3.
Pull off the mirror sleeve.