23

ACTION TAKEN

1.  

Check that the Main Power Switch is in the ON (I) position, and that the detachable power cord is  fully seated in the  

 

 

receptacle on back of system.  

2. 

Check that the system’s power cord plug is fully seated in an appropriate AC wall outlet. 

3. 

Check that the wall outlet is functional.

1.  

If the office has more than one Tap-On

TM

 Foot Pedal, test each to ensure that the proper Tap-On

TM

 Foot Pedal is being 

used.  With a handpiece and insert installed, depress the Tap-On

TM

 Foot Pedal to the first position.  The system should  

dispense water.  If none of the Tap-On

TM

 Foot Pedals operate the system, continue to the next step.

2. 

Resynchronize one Tap-On

TM

 Foot Pedal to the system (see Section 7.10 Tap-On

TM

 Foot Pedal Synchronization). 

1. 

Assure that handpiece lavage control is properly adjusted.

2. 

Check for clogged insert. Replace insert if necessary.

3. 

Check that dental office water supply valves are open.

4. 

If the system is connected to DualSelect Dispensing System, check that fluid level in the  selected bottle is sufficient.   

 

 

Make sure valves are open when using external water source.

5. 

Check that the water line filter is clean. Replace filter, if needed.

1. 

Check that the Power Control is not in Rinse Mode.

2. 

Check the insert for damage and that it is properly installed in the Handpiece.

3. 

Check that the handpiece is properly installed to the cable assembly.

4. 

Verify that the soft nozzle grip is flush with the hard plastic of the insert port. 

5. Turn the system’s Main Power Switch OFF, (O) position. Wait 5 seconds and turn the  system back ON.  

6. 

If problem still exists, replace both “AA” batteries in Tap-On

TM

 Foot Pedal with new “AA” batteries (Refer to Section 7.9) or 

connect Auxiliary  Tap-On

TM

 Foot Pedal Cable.

1. 

Fast blinking (3 blinks per second) – Indicates improper set-up

A.        If insert is in the handpiece, remove.  Verify the handpiece is properly seated and hold the Tap-On

TM

 Foot Pedal for 

2 seconds.   If blinking stops, the system is ready for use.  If blinking remains, continue to the next step.

B.        Attach a NEW handpiece and depress Tap-On

TM

 Foot Pedal for 2 seconds.  If blinking stops, the system is ready 

for use. Discard the  old handpiece or return if within warranty.  If blinking remains, continue to the next step.

C.        Install and fully seat an insert into handpiece.  Depress Tap-On

TM

 Foot Pedal for 2 seconds.  If blinking stops, the 

system is ready for  operation.  If blinking remains, continue to the next step.

D.        Install and fully seat a NEW insert in handpiece and depress Tap-On

TM

 Foot Pedal for 2 seconds.  If blinking stops, 

system is ready  for use.  Discard old insert or return if within warranty.  If blinking remains, refer to Section 11.2 

Technical Support and Repairs to have unit serviced as soon as possible.

2. 

Slow blinking (1 blink per second) - The system is not operating within factory specifications.

 

A. 

Remove insert.

 

B. 

Turn Main Power Switch OFF, (O) position.  Wait five seconds.  Turn unit ON, (I) position.

 

C. 

Operate Purge function.

 

D. 

If service indicator still blinks, refer to Section 11.2 Technical Support and Repairs to have unit serviced as soon    

  as 

possible.

1. 

Ensure that the base unit has adequate ventilation and is not near a heat source (i.e. radiator,  heat lamp, sunlight or  other  

 

heat producing operatory equipment).

2. 

Turn Main Power Switch OFF (O) position.  Allow system to cool for 10 minutes and turn system to the ON (I) position.     

 

Verify  light is not illuminated.

3. 

If light is still illuminated, refer to Section 11.2 Technical Support and Repairs to have unit serviced as soon as possible.

1. 

Check that there is no insert in the handpiece. 

2. 

Check that Handpiece is properly installed to the cable assembly.

QUICK REFERENCE GUIDE: TROUBLEShOOTING

SymPTOm

System will not 

operate: No Power  

ON Indicator

System will not 

operate: Power ON 

Indicator is illuminated 

 

System operates:  

No water flow  

to insert tip or 

handpiece  

overheats

System operates:  

No insert 

cavitation

System operates:

Service

indicator

blinking

System operates: 

Service indicator 

illuminated

System operates: 

Purge mode will not 

function -

 

icon flashing

Содержание Cavitron Plus

Страница 1: ...n TM Tap On TM tecnolog a Tap On TM tecnologia Tap On TM Tap On TM Technologie Technologie Tap On TM Tap On TM tecnolog a Tap On TM tecnologia Tap On TM Cavitron Plus Ultrasonic Scaler Please read car...

Страница 2: ...p On Technology does not function unless an insert is in the handpiece TIP Handpiece is detachable and autoclavable Make sure the electrical connections are dry before connecting Place ULTRASONIC INSE...

Страница 3: ...e foot pedal all the way to the floor enables quick removal of tenacious calculus Tip Quick release of pedal from Boost will not deactivate Tap On Technology Purge Remove insert from Handpiece and pre...

Страница 4: ...E Illuminates when the system is not functioning properly This display has three distinct modes Slow blink 1 blink per second means the system is not operating within factory specifications Fast blink...

Страница 5: ...gnostic Display Indicators and Controls 12 8 3 Handpiece Cable 13 8 4 Cavitron 30K Ultrasonic Inserts 13 8 5 Tap OnTM Wireless Foot Pedal Operation 14 8 6 Accessories and User Replaceable Parts 14 8 6...

Страница 6: ...e a Tap OnTM Wireless Foot Pedal with Tap On Technology Turbo Mode illuminated diagnostic display rinse setting automated purge function Steri Mate detachable sterilizable handpiece and 330 swivel han...

Страница 7: ...0601 Medical Device Standards Insufficient water flow could result in elevated water and tip temperature When operated at the input water temperature specified in the Water Line Requirements Section 7...

Страница 8: ...itron Plus Ultrasonic Scaler will create aerosols Following the procedural guidelines in Section 9 of this manual can effectively control and minimize aerosol dispersion 6 2 Water Supply Recommendatio...

Страница 9: ...it in direct sunlight may discolor plastic housing The system has been equipped with a Tap OnTM Wireless Foot Pedal which was factory synchronized to operate with the system s base unit If your office...

Страница 10: ...the following steps to synchronize the Tap OnTM Foot Pedal with the base unit 1 Turn the MainPower switch located at the center front underside of the system to the OFF O position 2 Install a new set...

Страница 11: ...n subgingivally scaling Rinse Turn the ultrasonic power level control knob fully counterclockwise until a click is heard Rinse mode is for use during an ultrasonic scaling procedure when lavage is des...

Страница 12: ...cator Illuminates when the Boost Mode has been activated with the Tap OnTM Foot Pedal Turbo Mode When pressed extra power is delivered to the system up to 25 This will increase the stroke of the tip T...

Страница 13: ...comfortable grasp of the handpiece The grip is sterilizable and is available in several different colors as an accessory for your Steri Mate Handpiece See installation instructions provided with the g...

Страница 14: ...chnology eliminates the need to hold the pedal down Tapping the foot pedal once activates ultrasonic power or rinse mode for approximately 4 minutes Tapping the foot pedal while in Tap On TM mode disa...

Страница 15: ...avitron Ultrasonic Inserts are intentionally rounded so there is minimal danger of tissue laceration with proper ultrasonic scaling technique Whenever the insert tip is placed in the patient s mouth t...

Страница 16: ...ect the handpiece cable for any breaks or tears 7 If using a closed water supply or DualSelect Dispensing System check for adequate fluid volumes for the next patient 8 When ready to use place a steri...

Страница 17: ...Tap OnTM Foot Pedal to the first position The system should dispense water If none of the Tap OnTM Foot Pedals operate the system continue to the next step 2 Resynchronize one Tap OnTM Foot Pedal to t...

Страница 18: ...ry Certified Service at 1 800 989 8826 Monday through Friday 8 00 A M to 5 00 P M Eastern Time For areas outside the U S contact your local DENTSPLY Professional representative Section 12 Warranty Per...

Страница 19: ...tion of ingress of water Consult Instructions for Use O I AC Power Switch 0 Off On Footswitch MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60...

Страница 20: ...61000 3 2 Class A Not Applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not Applicable The Ultrasonic Scaler is suitable for use in all establishments including domestic establishments a...

Страница 21: ...ite surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the high...

Страница 22: ...ectromagnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distance between Portable and mobile RF communications equipment and the mo...

Страница 23: ...blinking stops the system is ready for use If blinking remains continue to the next step B Attach a NEW handpiece and depress Tap OnTM Foot Pedal for 2 seconds If blinking stops the system is ready f...

Страница 24: ...riv e d air Installez les PILES dans la p dale sans fil Ins rez l insert ultrasonique dans la pi ce main R glez le niveau de puissance R glez la commande de lavage sur la pi ce main au niveau d sir Co...

Страница 25: ...cette fonction mains libres permet d augmenter la puissance du syst me pour retirer rapidement les d p ts tenaces il suffit d appuyer sur la p dale jusqu au bout vers le sol Astuce Le mode Tap OnTM re...

Страница 26: ...et indicateur peut s allumer de trois fa ons Clignotement lent 1 clignotement par seconde lorsque le syst me ne fonctionne pas selon les param tres d finis en usine Clignotement rapide 3 clignotements...

Страница 27: ...me 34 8 2 Voyants et commandes du tableau de diagnostic 35 8 3 Pi ce main c ble 36 8 4 Inserts ultrasoniques Cavitron 30K 36 8 5 Fonctionnement de la p dale Tap OnTM 37 8 6 Accessoires et pi ces remp...

Страница 28: ...Zone bleue laquelle permet une plus grande pr cision dans les param tres d alimentation Les fonctionnalit s volu es qui font du Cavitron Plus un investissement judicieux sont la p dale sans fil Tap O...

Страница 29: ...e sur un stimulateur cardiaque n ait t rapport DENTSPLY nous recommandons que les pi ces main et les c bles soient maintenus une distance d au moins 15 23 cm de tout appareil implant et de ses fils pe...

Страница 30: ...dicateur d efficacit d insert DENTSPLY Professional est inclus pour la commodit de l utilisateur Si une usure excessive est observ e si l insert a t pli ou d form ou s il pr sente d autres dommages je...

Страница 31: ...mati re d lectricit L alimentation lectrique du syst me doit amener un courant secteur monophas sous 100 240 volts de 50 60 Hz pouvant atteindre 1 0 amp re L alimentation lectrique doit arriver par l...

Страница 32: ...nTM Retournez la p dale sans fil Tap On et enlevez doucement la vis du couvercle du compartiment des piles l aide d un tournevis pointe cruciforme puis retirez le couvercle S il y a lieu enlevez les p...

Страница 33: ...es sont allum s appuyez sur le bouton Purge lui m me situ sur le tableau de diagnostic Les voyants se mettront clignoter dans un ordre s quentiel qui refl te le mode de synchronisation Cette phase dur...

Страница 34: ...om tre dans le sens inverse des aiguilles d une montre jusqu ce que vous entendiez un clic Le mode de rin age est utile durant une proc dure de d tartrage ultrasonique lorsque vous d sirez un lavage a...

Страница 35: ...S allume lorsque le mode Booster est activ l aide de la p dale Tap OnTM Mode Turbo Appuyez sur le bouton pour envoyer jusqu 25 de puissance en plus au syst me Cela allongera la course de l embout Le m...

Страница 36: ...lisable elle est disponible en plusieurs couleurs assorties votre pi ce main Steri Mate Veuillez consulter les instructions d installation fournies avec la poign e Les divers types d inserts ultrasoni...

Страница 37: ...e enfonc e Le fait de tapoter sur la p dale une fois active le mode de d tartrage ultrasons ou de rin age pendant environ 4 minutes Lorsque la p dale est en mode Tap On tapoter la p dale de nouveau po...

Страница 38: ...cadran et la surface en cours de traitement vacuez l eau d irrigation au moyen soit d un jecteur de salive soit d un HVE High Volume Evacuator 9 5 Ex cution des proc dures de d tartrage par ultrasons...

Страница 39: ...nouveau ou appuyez sur la p dale Tap On 6 Lorsque vous avez termin le cycle de purge placez un insert Cavitron 30 kHz st rilis dans la pi ce main puis r glez la commande de puissance ultrasonique et l...

Страница 40: ...du tuyau rel chez la pression d eau en appuyant sur la pointe du raccord et laissez l eau se vidanger dans un r cipient appropri 3 Saisissez le raccord sur l un des deux c t s du disque filtrant et to...

Страница 41: ...n enlevez le Assurez vous que la pi ce main est correctement enfonc e et tenez la p dale appuy e pendant 2 secondes Si le clignotement s arr te le syst me est pr t l emploi Si le clignotement persiste...

Страница 42: ...tinue 100 240 V c a Courant 1 0 amp re maximum Phase Simple Fr quence 50 60 Hertz Pression d eau 20 40 psig 138 275 kPa Temp rature de l eau 25 C 77 F Pression d air 65 100 psig 448 600 kPa D bit d ea...

Страница 43: ...Consulter le mode d emploi O I Interrupteur de courant alternatif 0 teint Allum Interrupteur au pied quipement m dical concernant les risques de choc lectrique d in cendie et d anomalie m canique uniq...

Страница 44: ...de tension scintillements IEC 61000 3 3 Non applicable Le d tartreur ultra sons peut tre utilis dans tous les types de b timents notamment les constructions domestiques et celles qui sont reli es dire...

Страница 45: ...ences b Des interf rences sont possibles aux alentours de l appareil elles sont identifi es avec ce symbole REMARQUE 1 A 80 MHz et 800 MHz la gamme de fr quence sup rieure s applique REMARQUE 2 Il se...

Страница 46: ...metteurs dont la puissance maximum en sortie n est pas mentionn e ci haut la distance de s paration d recommand e en m tres m peut tre estim e l aide de l quation applicable la fr quence de l metteur...

Страница 47: ...tape suivante B Branchez une pi ce main NEUVE et appuyez sur la p dale Tap On pendant 2 secondes Si le clignotement s arr te le syst me est pr t l emploi Jetez l ancienne pi ce main ou retournez la au...

Страница 48: ...800 989 8826 or 717 767 8502 De utschland France Australia DENTSPLY DeTrey GmbH DENTSPLY DeTrey DENTSPLY Australia Pty Ltd De Trey Strasse 1 17 Michael FARADAY 11 21 Gilby Road 78467 Konstanz 78380 M...

Отзывы: