FISDR-JAP / F1902124.X / 00 / 2011 – created 07/2011
- 31 / 208 -
7.
Close the battery cover.
8.
Tighten the screw with a screwdriver. Do not apply excessive force when tightening.
9.
Charge the battery before use.
NOTE
The used Nickel Metal Hydride batteries are recyclable, but their disposal may sometimes not
be permitted by law in your location. Return them to your distributor.
6.11.2.
Lubricating the Contra-angle
•
Lubricate the contra-angle only. We recommend using the X-SMART™ Spray.
•
Lubricate after each use and before sterilization.
1.
Screw the spray nozzle onto the spray with approx. 10 turns.
2.
Insert the spray nozzle into the rear part of the contra-angle and lubricate for 2-3 seconds
until oil comes out from the contra-angle head.
3.
Before attaching the lubricated contra-angle to the motor handpiece, wipe off excess oil.
Place it on its end or lean it at an angle for gravity draining. Mount it after excess oil has
been drained.
WARNING
Do not lubricate the motor handpiece.
CAUTIONS
•
Hold the contra-angle securely to prevent it from detaching with the pressure of the spray.
•
Never use a spray can upside down. Only spray gas, and no oil, would be sprayed.
.
F type Spray Nozzle
Contra-angle