Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
± 2 kV for power
supply lines
± 1 kV for input/output
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5%
U
T
(> 95% dip in
U
T
)
for 0,5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec
< 5%
U
T
(> 95% dip in
U
T
)
for 0,5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec
Mains power quality should be that of
a typical commercial or hospital
environment.
If the user of the device requires
continued operation during power
mains interruptions, it is recommended
that the device be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание X-SMART Plus
Страница 1: ...USER MANUAL...
Страница 3: ...FISDR JAP F1902124 X 00 2011 created 07 2011 4 208...
Страница 83: ...FISDR JAP F1902124 X 00 2011 created 07 2011 84 208...
Страница 163: ...FISDR JAP F1902124 X 00 2011 created 07 2011 164 208...
Страница 207: ...FISDR JAP F1902124 X 00 2011 created 07 2011 208 208...
Страница 208: ...FISDR JAP F1902124 X 00 2011 created 07 2011 209 208...