dentronix DDUS60 Скачать руководство пользователя страница 3

Owners Guide

3

Dentronix

®

Ultrasonic Cleaner

CAUTION

: Federal law restricts this device to sale by or on the

order of a dentist and/or physician.

An essential part of your infection control system

Congratulations on your purchase of a Dentronix Ultrasonic Cleaner.
Developed specifically for the orthodontic industry, Dentronix units
feature unique product design and optimize time and safety in the
office. Dentronix offers a complete line of ultrasonic cleaning sys-
tems and cleaning solutions which provide you with the flexibility to
efficiently clean instruments.

What is ultrasonic cleaning?

Ultrasonic Cleaning is a process created by high frequency sound
waves. The sound waves, enhanced by specially formulated clean-
ing solutions, create high energy cavitation. During cavitation,
millions of tiny bubbles form and then collapse or “implode,”
releasing enormous amounts of energy and shock waves which
scour the surface of instruments, appliances and other apparatus
placed in the cleaning solution. This powerful scouring action
reaches into minute crevices, which manual brushing cannot reach.
The combination of energy and specially formulated solutions make
ultrasonic cleaning the most effective method for removing gross
and microscopic debris.

I. Unpacking 

A. Remove your Dentronix unit from the shipping container

and

inspect the unit for any possible damage which may have
occurred during shipping. 

B. Check the serial number 

of the unit (found on the back side of

the unit) with the number on the shipping carton. If they are not
the same contact Dentronix.

C. Fill out the warranty card 

and mail to Dentronix within 10 days

of receipt.

D. Select a location 

for the cleaner. It should be close to a sink or

to waste lines to facilitate draining. The hose supplied with the
unit will allow you to place the unit on either side of the sink or
connect directly to the waste line. When used in conjunction with
a Dentronix DDS5000 dry heat sterilizer, find the best position for
your ultrasonic unit by calling Dentronix and asking for informa-
tion regarding “Sterilization Perfect Design.”

E. Connect one end of the hose to the drain valve (Fig. 1) 

and

place the other end in the sink. If desired (and recommended for
the DDUS60R), have your plumber connect the hose directly to
the waste system with the appropriate fittings per your local
building code.

II. Installation Instructions for the

DDUS60R Recessed Unit

The Dentronix DDUS60R consists of:

• Tank enclosure
• Electronics enclosure
• Drain tube
• Coaxial cable which is connected between the N female

coaxial connectors on the electronics control housing and the
tank housing

• Four-conductor cable which connects between DIN connec-

tors on the tank enclosure and the electronics enclosure

• Power supply cord

• Four (4) #8 Philips head wood screws

NOTE:

Before you begin, please read and understand the safety

instructions provided by the manufacturers of the tools required

for installation.

Tools Required

1 Electric Drill

1 Drill bit to drill a hole large enough to accept a reciprocating

saw blade

1 Reciprocating saw and blade

1 Phillips head screw driver

1 Flat blade screw driver

NOTE:

This Ultrasonic Cleaner must be installed by an experi-

enced equipment manager or licensed contractor. Be sure to

follow local building codes.

A. Tank Enclosure

1. Select a location for the cleaner. There must be at least 13"

clear vertical space below the top of the counter to accom-

modate the tank enclosure, and an additional 6" at the loca-

tion of the valve, for a total of 19". The countertop must be flat

and level. Make sure that you are not selecting a location

which will interfere with the drawers or vertical partitions

between cabinets.

2. Tape the enclosed template down to the countertop in the

desired location. Make sure the printed Dentronix logo on the

bezel faces you.

3. Drill four (4) 3/8" diameter holes where indicated on the

template (marked “A”).

4. Drill four (4) holes holes at the four locations marked “B.”

Select your hole size based upon the saw you will use in #5

below.

5. Cut along the “cut line” with a jigsaw or other reciprocating

saw.

6. Clean away dirt or sawdust around the opening, and remove

scrap wood.

7. Place tank enclosure into opening with the printed Dentronix

logo on the bezel facing you.

8. Assemble the bearing plates, lock washers and wing nuts to

the tie-down studs (per Detail A on the template).

9. Tighten all the wing nuts a few turns at a time until the bezel

just contacts the countertop. Once the bezel has contacted

the countertop on four sides, back off each wing nut 1/4 turn.

10.  Apply a thin bead of clear RTV silicone sealant between the

bezel and countertop.

B. Electronic Enclosure

1. Select a location for the electronics enclosure which will be

conveniently close to the tank enclosure.

NOTE:

Mount control unit so that the fuse drawer is acces-

sible (Section X, sub C). Be sure to leave at least 4" between

the back and sides of the electronic control housing and

cabinet wall. This will allow for adequate ventilation as well as

facilitate the insertion and removal of power/RF cables. 