VENTING
Standard mounting of the exhaust stack positions the opening of the stack pointing up (see Figure
3), or to the left when facing the back of the unit. CAUTION: Never point the stack opening down
towards the fan (unless a DAD-UL vent adapter is used with a flexible metal duct). To mount the
exhaust stack remove the three screws corresponding to the exhaust stack mounting holes in the back
of the sterilizer surrounding the exhaust opening (the opening with the exposed louvers). Place the
stack in position and secure with screws.
CAUTION
CAUTION: Do not remove cooling duct for any reason (See Figure 3)
Venting is recommended under the following conditions:
a) Installation of the DDS 7000 in a small room less than 12’ x 12’
(3.66m x 3.66m).
b) Installation of the DDS 7000 under cabinets or built into cabinets.
Two styles of vent adapters are available: (See Figure 4 Below)
Type DAD-UL – attaches to the exhaust stack allowing additional venting options – converts 5” x 7”
(12.7cm x 17.8cm) rectangular exhaust port to a 4” (10.2cm) diameter round flange.
Type DAD-BL – replaces the exhaust stack allowing vent connection directly to back of unit.
Both adapters utilize a 4” (10.2cm) diameter round duct. Metal or foil ducting material must be used. Vent
adapters can be used on vent runs of up to 6-8 ft (1.8m-2.4m). If your vent pipe is longer, sufficient
back pressure will be created in the airflow path reducing your unit’s cooling efficiency, resulting in extended
cycle times and possible E-19 error codes.
Figure 4
Figure 3
Exhaust Air
Cooling Duct
DAD-BL
DAD-UL