Pro-Select Platinum®
Operation and Instruction Manual
12
5. To reset default press
and hold until four beeps are heard
6. A light pen grasp using a stabilizing fulcrum, rather than a power fulcrum, is all that is required using PSP®
Irrigation mode
Attention
• Attach irrigator handpiece while in scaler mode or when power is off.
• Flush distilled water for 30 seconds when finished using medicaments.
1. Place patient in a more upright position
2. Press
to turn on PSP® or press
3. Flush irrigator handpiece with water in scaler mode
4. Select irrigator mode by using foot control or pressing
on control panel. LCD will show IRR and LCD will
blink while in irrigator mode. Irrigator handpiece will automatically begin a 45 second pre-heat cycle, when pre-
heat cycle is complete a beep will sound
5. Place cannula on irrigator handpiece
6. While waiting for heating cycle activation bend cannula as appropriate for access
7. Select desired fluid bottle
Use
or press
on control unit. A double beep will be heard as you change bottles. An LED light will
illuminate to indicate active bottle and LCD will show which bottle is active
8. Thorough low volume vacuum is sufficient for fluid evacuation from oral cavity
9. Prolonged inactivity will de-activate heater, you must ‘reboot’ by pressing
off, then
again
to re-activate 45-second pre-heat cycle.
10. Remove and properly discard cannula.
11. Flush irrigator handpiece with distilled water before autoclaving
12. Use
or press
to switch back to scaler mode. LCD will display scaler mode and current power level.
LCD will stop blinking
Attention
• Flow rate of liquid from irrigator handpiece is pre-set and not adjustable.
• Studies have shown heated chlorhexidine is more effective than at room
temperature*.
*Data on file