9
Pro-Select Platinum®
Operation and Instruction Manual
3. To connect scaler tip:
• Place handpiece in center of tip grip
• Using a clockwise rotation, screw tip onto handpiece
• Align tip wrench with flat sides of tip base
• Use tip grip and wrench to tighten until tip is secure
4. To remove the scaler tip:
• Place handpiece in center of tip grip
• Turn wrench counterclockwise to loosen tip and remove tip with fingers
Connect irrigator tip
Warning
Irrigation handpiece heater has been designed for use with a 23
gauge side ported cannula. Do not use other gauge cannulas, as
fluid heat variation will result.
1. Slide 23 gauge side ported cannula on irrigation handpiece
2. Using cannula cap, bend cannula for proper adaptation
Attention
An irrigator cannula is a single patient use, disposable item
and should NOT be sterilized.
Connect handpiece to cable
Attention
LED will remain activated for 10 seconds after foot control is released.
1. Plug handpiece into cable
2. Place handpiece in cradle when not in use
!