Resistance to electromagnetic fields
Test
Test level
Level of
compliance
Terms of use
Electrostatic
discharge
(ESD) IEC
61000-4-2
±6 kV
contact ±8
kV impact
±4 kV
contact ±8
kV impact
Floor should be made from wood, concrete, or
ceramic tiles. If floor is covered with synthetic
material, the relative humidity should be at
least 40%.
Magnetic
fields
IEC 610004-8
3 A/m
3 A/m
Parameters of magnetic fields should be
at levels, typical for commercial or hospital
buildings