Demtech Pro-Tester SE Скачать руководство пользователя страница 13

Rev. B, October 2020 

Pro-Tester SE Operator’s Manual   

 

 

 

 

 

                Page 10 

 
 

5.5

 

Settings 

 

a.

 

The Pro-Tester

TM

 requires various settings for proper operation.  These settings are 

easily accessed via the Menu Mode of operation. To enter the Menu Mode, make sure 
the Pro-Tester

TM

 is on and at the main operating screen (Refer to Page 15).  

Simultaneously 

press and hold the YES and NO

 buttons on the Control Panel for 

several seconds. This procedure puts the system into Menu Mode.  This mode of 
operation allows the user the set various system parameters.  For a complete listing of 
all of the features refer to the diagram on page 15.  

b.

 

All of the settings required to operate the Pro-Tester

TM

 come factory set for most users 

and should not need setting. However, if your particular installation requires setting 
any parameters enter the Menu Mode and follow the screen prompts to navigate the 
screen menu. 

c.

 

Figure 7.1 on page 15 is a Menu Tree drawing. This drawing can be very helpful if 
you are trying to find a particular function or setting you wish to access.  Use the 
drawing to locate the particular item you want to access.  Once located, follow the 
tree back to the top level.  Next, utilizing the user interface on the Pro-Tester

TM

proceed back down through the menu structure using the drawing as a guide to access 
the particular screen you want. 

 

5.6

 

Elongation Definition 

 

If you intend to use the strain (elongation) feature of the Pro-Tester

TM

 tensiometer it is 

important to understand both the definition of this feature/function and how it is intended 
to be used. Proper understanding requires the definition of some basic terms.  These 
terms are defined below. 

 
 

a.

 

Grip Separation (GS

): The distance between the grip point on each of the left and 

right sample grip assemblies (refer to Figure 5.5).  There are two user settable 
distances, A or B, which have been factory set for 1.7 and 3.5 inches respectively.  
The operator must be aware of these values and change the settings if necessary (refer 
to Section 5.5) to accommodate the particular sample to be tested. 

b.

 

Weld Extrusion (Squeeze-out):

 The area of a weld whereupon becoming molten the 

parent material extrudes (is squeezed) out from between the pressure zone of the 
welding tool and then cools.  This extruded material does not usually provide an 
acceptable bond of the parent material and should not be considered part of the weld 
(refer to Figure 6C). 

Содержание Pro-Tester SE

Страница 1: ...Pro Tester SE Pro Tester SE Operator s Manual Model Number T 0100SE A Revision B...

Страница 2: ...mation retrieval system or otherwise without prior written permission of DEMTECH Services Inc Notice Patents have been granted and or patent applications are pending or are in the process of preparati...

Страница 3: ...PRODUCT DATA PAGE 3 4 SITE PREPARATION PAGE 4 5 UNPACKING INSTALLATION INSTRUCTIONS PAGE 4 5 1 Unpacking and Preparation Page 5 5 2 Electrical Plug Connection Page 5 5 3 Control Panel Page 7 5 4 Load...

Страница 4: ...time while it is plugged in and the power is turned on Incorrect use of the product such as overheating of the material can present a fire and explosion hazard especially near combustible materials an...

Страница 5: ...tions However improper use or abuse may lead to hazardous conditions for the user or other personnel or cause damage to the tensiometer Always keep this manual with the Pro Tester TM at or near the lo...

Страница 6: ...utton operation combined with automatic return saves time The enclosed controller system is easy to maintain You can be assured the Pro Tester TM has been built using the highest quality materials ava...

Страница 7: ...tailed unpacking and installation instructions provided in this manual 5 Unpacking Installation Instructions This section contains instructions for the unpacking placement and installation of the Pro...

Страница 8: ...d injury and or damage to the system be sure to use adequate care when lifting removing or replacing the system into the shipping transport container e You may operate the system from within the shipp...

Страница 9: ...the grip head in left to find the home position then out right to calibrate travel speed After a short pause the grip will then move back in left and stop at the A or B grip separation position depend...

Страница 10: ...Operator s Manual 5 3 Control Panel The User Interface Control Panel is the main operator interface device for operating the Pro Tester TM Each component of the Control Panel is identified below refe...

Страница 11: ...and is used to change various settings during Menu Mode operation refer to Section 6 5 c Grip Start Position Toggle Switch A B This switch allows the user to select between one of two grip separation...

Страница 12: ...unscrewing the tri handle counter clockwise open The sample is prepared by bending the ends to be clamped so they can be easily loaded into the grips Load the sample from the top of the grip until the...

Страница 13: ...want to access Once located follow the tree back to the top level Next utilizing the user interface on the Pro Tester TM proceed back down through the menu structure using the drawing as a guide to a...

Страница 14: ...is test d Gauge Length GL This is the base value used by the Pro Tester TM for determining elongation percentage It is the Grip Separation once the initial tension value has been reached minus the Wel...

Страница 15: ...Rev B October 2020 Pro Tester SE Operator s Manual Page 12 Figure 6A Figure 6B Figure 6C Figure 6...

Страница 16: ...H Services Inc at DemTech Services Inc Ship to address 6414 Capitol Avenue Diamond Springs CA 95619 U S A 7 User Interface Menu Tree Refer to Page 15 for the Pro Tester SE TM Menu Tree drawing This dr...

Страница 17: ...Refer to Page 16 for the Pro Tester TM System Wire Diagram This drawing can be very helpful in identifying and or troubleshooting basic wiring or interconnection issues related to the Pro Tester TM 9...

Страница 18: ...Rev B October 2020 Page 15 Pro Tester SE Operator s Manual Figure 7 1 T 0100HP A M001...

Страница 19: ...Rev B October 2020 Pro Tester SE Operator s Manual Page 16 Figure 8 1...

Страница 20: ...subjected to abuse accident or other abnormal conditions IF ANY FAULT DEVELOPS THE FOLLOWING STEPS SHOULD BE TAKEN 1 Notify DEMTECH Services Inc by calling 1 888 324 9353 Overseas customers should co...

Отзывы: