Alarm delay can be programmed to trigger only when temperature
excursions continue beyond the set delay period that starts with
the first alarm event. The buzzer can be temporarily silenced by pressing the
STOP
button and will restart when the next temperature reading is recorded.
Note: The buzzer consumes more energy and should be stopped as soon
as possible to conserve battery life.
Level 1 High/Low Alarm
Numeral 1 icon
Yellow LED flashes every 5 seconds
High or Low
Buzzer sound 1 beep at set interval
Level 2 Extreme High/Low Alarm
Numeral 2 icon
Red LED flashes every 5 seconds
High or Low
Buzzer sound 3 beeps at set interval
The LCD Screen
The Glycol Bottle
This digital data logger (DDL) features a stainless steel sensor enclosed in a
bottle of propylene glycol. The cable has a stereo jack for quick disconnect
to easily detach the logger and download data. The glycol buffered sensor
emulates products in the storage unit and keeps the sensor from setting off
false alarms due to rapid temperature fluctuations caused by frequent door
opening during normal daily use.
Place the glycol bottle sensor in the center of the refrigerator or freezer, attach
1
11
5 6
4
3
10
2
12
8
9
S
°C
°F
%
MAX
MIN
AVG
HI
LO
day
hrs
min
7
1.
Sensor/alarm channel enumerator
2.
Shadow logging icon
3.
Active mode icon
4.
Stop mode icon
5.
High and low alarm arrows
6.
Alarm icon
7.
°C, °F & % scale icons
8.
Day, Hours and Minutes icons
9.
High and Low Alarm set point icons
10.
Statistical icons
11.
4 digit numerical display
12.
Battery Capacity Gauge
°C
HI
°C
LO