![DeltaOHM HD2402 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/deltaohm/hd2402/hd2402_operating-manual_2479058047.webp)
HD2402
- 47 -
V1.4
7.2
S
PECTRAL INTERVAL
315-400nm (UVA)
No weighting is required by this spectral interval.
The following figure shows the spectral response curve of the UVA channel.
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
250
270
290
310
330
350
370
390
410
430
450
Lambda(nm)
Figure 8: spectral response curve
of
UVA sensor
7.3
S
PECTRAL INTERVAL
300-700nm
This spectral interval must be evaluated with the weighting curve B(
λ
).
The following figure shows the comparison between the spectral response curve ob-
tained by the combination of channels and weighting curve B(
λ
).
1.0E-05
1.0E-04
1.0E-03
1.0E-02
1.0E-01
1.0E+00
300
350
400
450
500
550
600
650
700
Lambda(nm)
Figure 9
:
comparison between spectral response curve obtained by the combina-
tion of channels (in purple color) and weighting curve B(
λ
) (in black)