HD2402
- 39 -
V1.4
it is assumed that C
α
=100
if L
r_100
< 6
•
10
4
⇒
t_lim > 10s (j limit satisfied)
if 2.8
•
10
5
≤
L
r_100
≤
8.89
•
10
6
⇒
t_lim = (5
•
10
5
/L
r_100
)
4
(k limit satisfied)
if L
r_100
< 8.89
•
10
6
⇒
t_lim < 10
μ
s (l limit)
o
L
r
(11mrad) [W/(m² sr)] radiance calculated assuming that all the light comes
from a circular area whose angular dimension is 11mrad. This value is to be
compared with the three limit values.
o
t_
lim11
[s] exposure limit time, that is the time period during which a source can
be observed without damages to the visual system. This value is calculated with
the following procedure (all radiance values are expressed in [W m
-2
sr
-1
]):
it is assumed that C
α
= 11
if L
R_11
≤
5.45
•
10
5
⇒
t_lim > 10s (j limit satisfied)
if 5.45
•
10
5
≤
L
R_11
≤
8.08
•
10
7
⇒
t_lim
=
(4.54
•
10
6
/L
R_11
)
4
(k limit satisfied)
if L
R_11
> 8.08
•
10
7
⇒
t_lim < 10
μ
s (l limit)
•
Alpha [rad]
, angular dimension of the source in radiants supplied by the user.
o
Alpha [rad] angle subtended by the source.
o
Omega [sr] solid angle subtended by the source, calculated starting from Alpha
angle, assuming that the source shape is circular:
Omega = 2
π
(1-cos(Alpha/2)
o
L
R
(100mrad) [W/(m² sr)] radiance calculated assuming that the source angular
dimension is 100mrad.
o
L
R
(11mrad) [W/(m² sr)] radiance calculated assuming that the source angular
dimension is 11mrad.
o
L
R
(real) [W/(m² sr)] is the real radiance calculated starting from the entered
geometric parameters.
o
t_lim [s] exposure limit time, that is the time period during which a source can
be observed without damages to the visual system. This value is calculated with
the following procedure (all radiance values are expressed in [W m
-2
sr
-1
]):
if Alpha > 100mrad it is assumed that C
α
= 100
if L
R_real
≤
6
•
10
4
⇒
t > 10s (j limit satisfied)
if 6
•
10
4
≤
L
R_real
≤
8.89
•
10
6
⇒
t_lim = (5
•
10
5
/L
R_real
)
4
[s] (k limit satisfied)
if 8.89
•
10
6
≤
L
R_real
⇒
t_lim < 10
μ
s (l limit)
if 11mrad < Alpha < 100mrad it is assumed that C
α
= Alpha
if L
R_real
≤
6
•
10
6
/Alpha
⇒
t > 10s (j limit satisfied)
if 6
•
10
6
/Alpha
≤
L
R_real
≤
8.89
•
10
8
/Alpha
⇒
t_lim = (5
•
10
7
/(L
R_real
•
Alpha))
4
[s]
(k limit satisfied)
if 8.89
•
10
8
/Alpha
≤
L
R_real
⇒
t_lim < 10
μ
s (l limit)